Recent advances in artificial intelligence have enabled promising applications in neurosurgery that can enhance patient outcomes and minimize risks. This paper presents a novel system that utilizes AI to aid neurosurgeons in precisely identifying and localizing brain tumors. The system was trained on a dataset of brain MRI scans and utilized deep learning algorithms for segmentation and classification. Evaluation of the system on a separate set of brain MRI scans demonstrated an average Dice similarity coefficient of 0.87. The system was also evaluated through a user experience test involving the Department of Neurosurgery at the University Hospital Ulm, with results showing significant improvements in accuracy, efficiency, and reduced cognitive load and stress levels. Additionally, the system has demonstrated adaptability to various surgical scenarios and provides personalized guidance to users. These findings indicate the potential for AI to enhance the quality of neurosurgical interventions and improve patient outcomes. Future work will explore integrating this system with robotic surgical tools for minimally invasive surgeries.
What might the attendee be able to do after being in your session?
Our work shows how to connect intra-operative devices via IEEE 11073 Service-oriented Device Connectivity (SDC).
Description of the Problem or Gap
Standardized device communication is essential for interoperability, availability of device data, and therefore for the intelligent operating room (OR) and arising solutions. The SDC standard was developed to make information from medical devices available in a uniform manner and enable interoperability. Existing devices are rarely SDC-capable and need interfaces to be interoperable via SDC.
Methods: What did you do to address the problem or gap?
We conceived an SDC-based architecture consisting of a service provider and service consumer. In our concept, the service provider is connected to the medical device and capable to translate the proprietary protocol of the device into SDC and vice versa. The service consumer is used to request or send information via the SDC protocol to the service provider and can function as a uniform bidirectional interface (e.g. for displaying or controlling). This concept was exemplarily demonstrated with the patient monitor MX800 of Philips to retrieve the device data (e.g. vital parameters) via SDC and partly for the operating light marLED X of KLS Martin Group.
Results: What was the outcome(s) of what you did to address the problem or gap?
The patient monitor MX800 was connected to a Raspberry Pi (RPi) via LAN, on which the service provider is running. The python script on the RPi establishes a connection to the monitor and translates incoming and outgoing messages from the proprietary protocol to SDC and vice versa to/from the service consumer. The service consumer is running on a laptop and acts as a simulation for different kinds of systems that want to get vital parameters or other information from the patient monitor. The operating light marLED X was connected to an RPi via USB-to-RS232. A python script on the RPi establishes a connection to the light and makes it possible via proprietary commands to get information of the light (e.g. status) and to control it (e.g. toggle the light, increment the intensity). A translation to SDC is not integrated yet.
Discussion of Results
Our practical implementation shows that medical devices can be accessed via external connections to get device data and control the device via commands. The example SDC implementation of the patient monitor MX800 makes it possible to request its data via the standardized communication protocol SDC. This is also possible for the operating light marLED X if its proprietary protocol is analyzed to be translatable to/from SDC. This would allow to control the device from an external system, or automatically depending on the status of the ongoing procedure. The advantage is, that existing intra-operative devices can be extended by a service provider which is capable of translating the proprietary protocol of the device in SDC and vice versa. This enables interoperability and an intelligent OR that, for example, is aware of all devices, their status, and data and can use this information to optimally support the surgeons and their team (e.g. provision of information, automated documentation). This interoperability allows that future innovations merely need to understand the SDC protocol instead of all vendor-dependent communication protocols.
Conclusion
Standardized device communication is essential to reach interoperability, and therefore intelligent ORs. Our contribution addresses the possibility of subsequently making medical devices SDC-capable. This may eliminate the need of understanding all the different proprietary protocols when developing new innovative solutions for the OR.
Introduction: Even if there is a standard procedure of CI surgery, especially in pediatric surgery surgical steps often differ individually due to anatomical variations, malformations or unforseen events. This is why every surgical report should be created individually, which takes time and relies on the correct memory of the surgeon. A standardized recording of intraoperative data and subsequent storage as well as text processing would therefore be desirable and provides the basis for subsequent data processing, e.g. in the context of research or quality assurance.
Method: In cooperation with Reutlingen University, we conducted a workflow analysis of the prototype of a semi-automatic checklist tool. Based on automatically generated checklists generated from BPMN models a prototype user interface was developed for an android tablet. Functions such as uploading photos and files, manual user entries, the interception of foreseeable deviations from the normal course of operations and the automatic creation of OP documentation could be implemented. The system was tested in a remote usability test on a petrous bone model.
Result: The user interface allows a simple intuitive handling, which can be well implemented in the intraoperative setting. Clinical data as well as surgical steps could be individually recorded and saved via DICOM. An automatic surgery report could be created and saved.
Summary: The use of a dynamic checklist tool facilitates the capture, storage and processing of surgical data. Further applications in clinical practice are pending.
This project aims to evaluate existing big data infrastructures for their applicability in the operating room to support medical staff with context-sensitive systems. Requirements for the system design were generated. The project compares different data mining technologies, interfaces, and software system infrastructures with a focus on their usefulness in the peri-operative setting. The lambda architecture was chosen for the proposed system design, which will provide data for both postoperative analysis and real-time support during surgery.
Purpose
For the modeling, execution, and control of complex, non-standardized intraoperative processes, a modeling language is needed that reflects the variability of interventions. As the established Business Process Model and Notation (BPMN) reaches its limits in terms of flexibility, the Case Management Model and Notation (CMMN) was considered as it addresses weakly structured processes.
Methods
To analyze the suitability of the modeling languages, BPMN and CMMN models of a Robot-Assisted Minimally Invasive Esophagectomy and Cochlea Implantation were derived and integrated into a situation recognition workflow. Test cases were used to contrast the differences and compare the advantages and disadvantages of the models concerning modeling, execution, and control. Furthermore, the impact on transferability was investigated.
Results
Compared to BPMN, CMMN allows flexibility for modeling intraoperative processes while remaining understandable. Although more effort and process knowledge are needed for execution and control within a situation recognition system, CMMN enables better transferability of the models and therefore the system. Concluding, CMMN should be chosen as a supplement to BPMN for flexible process parts that can only be covered insufficiently by BPMN, or otherwise as a replacement for the entire process.
Conclusion
CMMN offers the flexibility for variable, weakly structured process parts, and is thus suitable for surgical interventions. A combination of both notations could allow optimal use of their advantages and support the transferability of the situation recognition system.
Automatic segmentation is essential for the brain tumor diagnosis, disease prognosis, and follow-up therapy of patients with gliomas. Still, accurate detection of gliomas and their sub-regions in multimodal MRI is very challenging due to the variety of scanners and imaging protocols. Over the last years, the BraTS Challenge has provided a large number of multi-institutional MRI scans as a benchmark for glioma segmentation algorithms. This paper describes our contribution to the BraTS 2022 Continuous Evaluation challenge. We propose a new ensemble of multiple deep learning frameworks namely, DeepSeg, nnU-Net, and DeepSCAN for automatic glioma boundaries detection in pre-operative MRI. It is worth noting that our ensemble models took first place in the final evaluation on the BraTS testing dataset with Dice scores of 0.9294, 0.8788, and 0.8803, and Hausdorf distance of 5.23, 13.54, and 12.05, for the whole tumor, tumor core, and enhancing tumor, respectively. Furthermore, the proposed ensemble method ranked first in the final ranking on another unseen test dataset, namely Sub-Saharan Africa dataset, achieving mean Dice scores of 0.9737, 0.9593, and 0.9022, and HD95 of 2.66, 1.72, 3.32 for the whole tumor, tumor core, and enhancing tumor, respectively.
Motivation
In order to enable context-aware behavior of surgical assistance systems, the acquisition of various information about the current intraoperative situation is crucial. To achieve this, the complex task of situation recognition can be delegated to a specialized system. Consequently, a standardized interface is required for the seamless transfer of the recognized contextual information to the assistance systems, enabling them to adapt accordingly.
Methods
Our group analyzed four medical interface standards to determine their suitability for exchanging intraoperative contextual information. The assessment was based on a harmonized data and service model derived from the requirements of expected context-aware use cases. The Digital Imaging and Communications in Medicine (DICOM) and IEEE 11073 for Service-oriented Device Connectivity (SDC) were identified as the most appropriate standards.
Results
We specified how DICOM Unified Procedure Steps (UPS), can be used to effectively communicate contextual information. We proposed the inclusion of attributes to formalize different granularity levels of the surgical workflow.
Conclusions
DICOM UPS SOP classes can be used for the exchange of intraoperative contextual information between a situation recognition system and surgical assistance systems. This can pave the way for vendor-independent context awareness in the OR, leading to targeted assistance of the surgical team and an improvement of the surgical workflow.
Purpose
Artificial intelligence (AI), in particular deep learning (DL), has achieved remarkable results for medical image analysis in several applications. Yet the lack of human-like explanations of such systems is considered the principal restriction before utilizing these methods in clinical practice (Yang, Ye, & Xia, 2022).
Methods
Explainable Artificial Intelligence (XAI) provides a human-explainable and interpretable description of the “black-box” nature of DL (Gulum, Trombley, & Kantardzic, 2021). An effective XAI diagnosis generator, namely NeuroXAI (refer to Fig. 1), has been developed to extract 3D explanations from convolutional neural networks (CNN) models of brain gliomas (Zeineldin et al., 2022). By providing visual justification maps, NeuroXAI can help make DL models transparent and thus increase the trust of medical experts.
Results
NeuroXAI has been applied to two applications of the most widely investigated problems in brain imaging analysis, i.e. image classification and segmentation using magnetic resonance imaging (MRI). Visual attention maps of multiple XAI methods have been generated and compared for both applications, which could help to provide transparency about the performance of DL systems.
Conclusion
NeuroXAI helps to understand the prediction process of 3D CNN networks for brain glioma using human-understandable explanations. Results revealed that the investigated DL models behave in a logical human-like manner and can improve the analytical process of the MRI images systematically. Due to its open architecture, ease of implementation, and scalability to new XAI methods, NeuroXAI could be utilized to assist medical professionals in the detection and diagnosis of brain tumors. NeuroXAI code is publicly accessible at https://github.com/razeineldin/NeuroXAI
Purpose
Supporting the surgeon during surgery is one of the main goals of intelligent ORs. The OR-Pad project aims to optimize the information flow within the perioperative area. A shared information space should enable appropriate preparation and provision of relevant information at any time before, during, and after surgery.
Methods
Based on previous work on an interaction concept and system architecture for the sterile OR-Pad system, we designed a user interface for mobile and intraoperative (stationary) use, focusing on the most important functionalities like clear information provision to reduce information overload. The concepts were transferred into a high-fidelity prototype for demonstration purposes. The prototype was evaluated from different perspectives, including a usability study.
Results
The prototype’s central element is a timeline displaying all available case information chronologically, like radiological images, labor findings, or notes. This information space can be adapted for individual purposes (e.g., highlighting a tumor, filtering for own material). With the mobile and intraoperative mode of the system, relevant information can be added, preselected, viewed, and extended during the perioperative process. Overall, the evaluation showed good results and confirmed the vision of the information system.
Conclusion
The high-fidelity prototype of the information system OR-Pad focuses on supporting the surgeon via a timeline making all available case information accessible before, during, and after surgery. The information space can be personalized to enable targeted support. Further development is reasonable to optimize the approach and address missing or insufficient aspects, like the holding arm and sterility concept or new desired features.
Background
Personalized medicine requires the integration and analysis of vast amounts of patient data to realize individualized care. With Surgomics, we aim to facilitate personalized therapy recommendations in surgery by integration of intraoperative surgical data and their analysis with machine learning methods to leverage the potential of this data in analogy to Radiomics and Genomics.
Methods
We defined Surgomics as the entirety of surgomic features that are process characteristics of a surgical procedure automatically derived from multimodal intraoperative data to quantify processes in the operating room. In a multidisciplinary team we discussed potential data sources like endoscopic videos, vital sign monitoring, medical devices and instruments and respective surgomic features. Subsequently, an online questionnaire was sent to experts from surgery and (computer) science at multiple centers for rating the features’ clinical relevance and technical feasibility.
Results
In total, 52 surgomic features were identified and assigned to eight feature categories. Based on the expert survey (n = 66 participants) the feature category with the highest clinical relevance as rated by surgeons was “surgical skill and quality of performance” for morbidity and mortality (9.0 ± 1.3 on a numerical rating scale from 1 to 10) as well as for long-term (oncological) outcome (8.2 ± 1.8). The feature category with the highest feasibility to be automatically extracted as rated by (computer) scientists was “Instrument” (8.5 ± 1.7). Among the surgomic features ranked as most relevant in their respective category were “intraoperative adverse events”, “action performed with instruments”, “vital sign monitoring”, and “difficulty of surgery”.
Conclusion
Surgomics is a promising concept for the analysis of intraoperative data. Surgomics may be used together with preoperative features from clinical data and Radiomics to predict postoperative morbidity, mortality and long-term outcome, as well as to provide tailored feedback for surgeons.