Perforations of the tympanic membrane (TM) can occur as a result of injury or inflammation of the middle ear. These perforations can lead to conductive hearing loss (HL), where in some cases the magnitude of HL exceeds that attributable to the observed TM perforation alone. We aim with this study to better understand the effects of location and size of TM perforations on the sound transmitting properties of the middle ear.
The middle ear transfer function (METF) of six human temporal bones (TB; freshly frozen specimen of body donors) were compared before and after perforation of the TM at different locations (anterior or posterior lower quadrant) and of different sizes (1mm, ¼ of the TM, ½ of the TM, and full ablation). The
METF were correlated with a Finite Element (FE) model of the middle ear, in which similar alterations were simulated.
The measured and simulated FE model METFs exhibited frequency and perforation size dependent amplitude losses at all locations and severities. In direct comparison, posterior TM perforations affected the transmission properties to a larger degree than perforations of the anterior quadrant. This could possibly be caused by an asymmetry of the TM, where the malleus-incus complex rotates and results in larger deflections in the posterior TM half than in the anterior TM half. The FE model of the TM with a sealed cavity suggest that small perforations result in a decrease of TM rigidity and thus to an increase in oscillation amplitude of the TM, mostly above 1 kHz.
The location and size of TM perforations influence the METF in a reproducible way. Correlating our data with the FE model could help to better understand the pathologic mechanisms of middle-ear diseases. If small TM perforations with uncharacteristically significant HL are observed in daily clinical practice, additional middle ear pathologies should be considered. Further investigations on the loss of TM pretension due to perforations may be informative.
In order to evaluate the performance of different stapes prosthesis types, a coupled finite element (FE) model of human ear was developed. First, the middle-ear FE model was developed and validated using the middle-ear transfer function measurements available in literature including pathological cases. Then, the inner-ear FE model was developed and validated using tonotopy, impedance, and level of cochlea amplification curves from literature. Both models are based on pre-existing research with some improvements and were combined into one coupled FE model. The stapes in the coupled FE ear model was replaced with a model of a stapes prosthesis to create a reconstructed ear model that can be used to estimate how different types of protheses perform relative to each other as well as to the natural ear. This will help in designing of new innovative types of stapes prostheses or any other type of middle-ear prostheses as well as to improve the ones that are already available on the market.
Simulation models of the middle ear have rarely been used for diagnostic purposes due to their limited predictive ability with respect to pathologies. One big challenge is the large uncertainty and ambiguity in the choice of material parameters of the model.
Typically, the model parameters are determined by fitting simulation results to validation measurements. In a previous study, it was shown that fitting the model parameters of a finite-element model using the middle-ear transfer function and various other measurable output variables from normal ears alone is not sufficient to obtain a good predictive ability of the model on pathological middle-ear conditions. However, the inclusion of validation measurements on one pathological case resulted in a very good predictive ability also for other pathological cases. Although the found parameter set was plausible in all aspects, it was not yet possible to draw conclusions about the uniqueness and the accuracy or the uncertainty of the parameter set.
To answer these questions, statistical solution approaches are used in this study. Using the Monte Carlo method, a large number of plausible model data sets are generated that correctly represent the normal and pathological middle-ear characteristics in terms of various output variables like e.g., impedance, reflectance, umbo, and stapes transfer function. Subsequent principal component analyses (PCA) allow to draw conclusions about correlations, quantitative limits and statistical density of parameter values.
Furthermore, applying inverse PCA yields numerous plausible parameterizations of the middle-ear model, which can be used for data augmentation and training of a neural network which is capable of distinguishing between a normal middle ear and pathologies like otosclerosis, malleus fixation, and disarticulation based on objectively measured quantities like impedance, reflectance, and umbo velocity.
The hearing contact lens® (HCL) is a new type of hearing aid devices. One of its main components is a piezo-electric actuator. In order to evaluate and maximize the HCL’s performance, a model of the HCL coupled to the middle ear was developed using finite element approach. The model was validated step by step starting with the HCL only. To validate the HCL model, vibrational measurements on the HCL were performed using a Laser-Doppler-Vibrometer (LDV). Then, a silicone cap was placed onto the HCL to provide an interface between the HCL and the tympanic membrane of the middle-ear model and additional LDV measurements on temporal bones were performed to validate the coupled model. The coupled model was used to evaluate the equivalent sound pressure of the HCL. Moreover, a deeper insight was gained into the contact between the HCL and tympanic membrane and its effects on the HCL performance. The model can be used to investigate the sensitivity of geometrical and material parameters with respect to performance measures of the HCL and evaluate the feedback behavior.
The hearing contact lens® (HCL) is a new type of hearing aid devices. One of its main components is a piezo-electric actuator. In order to evaluate and maximize the HCL's performance, a model of the HCL coupled to a middle-ear model was developed using finite element approach. The model was validated step by step starting with the HCL only. To validate the HCL model, vibrational measurements on the HCL were performed using a laser-doppler-vibrometer (LDV). Then, a silicone cap was placed onto the HCL to provide an interface between the HCL and the tympanic membrane of the middle-ear model, and additional LDV measurements on temporal bones were performed to validate the coupled model that was used to evaluate the equivalent sound pressure of the HCL. Moreover, a de-eper insight was gained into the contact between the HCL and tympanic membrane and its effects on the HCL performance. The model can be used to investigate the sensitivity of geometrical and material parameters with respect to performance measures of the HCL and evaluate the feedback behavior.
Hearing contact lens (HCL) is a new type of hearing aid devices. One of its main components is a piezo-electric actuator (PEA). In order to evaluate and maximizethe HCL´s performance, a model of the HCL coupled to the middle ear was developed using finite element (FE)approach. To validate the model, vibrational measurements on the HCL and temporal bones were performed using a Laser-Doppler-Vibrometer (LDV). The model was validated step by step starting with HCL only. Then a silicone cap was fitted onto the HCL to provide an interface between the HCL and the tympanic membrane. The HCL was placed on the tympanic membrane and additional measurements were performed to validate the coupled model. The model was used to evaluate the sensitivity of geometrical and material parameters with respect to performance measures of the HCL. Moreover, deeper insight was gained into the feedback behavior, which causes whistling sounds, and the contact between the HCL and tympanic membrane.
Current clinical practice is often unable to identify the causes of conductive hearing loss in the middle ear with sufficient certainty without exploratory surgery. Besides the large uncertainties due to interindividual variances, only partially understood cause–effect principles are a major reason for the hesitant use of objective methods such as wideband tympanometry in diagnosis, despite their high sensitivity to pathological changes. For a better understanding of objective metrics of the middle ear, this study presents a model that can be used to reproduce characteristic changes in metrics of the middle ear by altering local physical model parameters linked to the anatomical causes of a pathology. A finite-element model is, therefore, fitted with an adaptive parameter identification algorithm to results of a temporal bone study with stepwise and systematically prepared pathologies. The fitted model is able to reproduce well the measured quantities reflectance, impedance, umbo and stapes transfer function for normal ears and ears with otosclerosis, malleus fixation, and disarticulation. In addition to a good representation of the characteristic influences of the pathologies in the measured quantities, a clear assignment of identified model parameters and pathologies consistent with previous studies is achieved. The identification results highlight the importance of the local stiffness and damping values in the middle ear for correct mapping of pathological characteristics and address the challenges of limited measurement data and wide parameter ranges from the literature. The great sensitivity of the model with respect to pathologies indicates a high potential for application in model-based diagnosis.
Purpose
Injury or inflammation of the middle ear often results in the persistent tympanic membrane (TM) perforations, leading to conductive hearing loss (HL). However, in some cases the magnitude of HL exceeds that attributable by the TM perforation alone. The aim of the study is to better understand the effects of location and size of TM perforations on the sound transmission properties of the middle ear.
Methods
The middle ear transfer functions (METF) of six human temporal bones (TB) were compared before and after perforating the TM at different locations (anterior or posterior lower quadrant) and to different degrees (1 mm, ¼ of the TM, ½ of the TM, and full ablation). The sound-induced velocity of the stapes footplate was measured using single-point laser-Doppler-vibrometry (LDV). The METF were correlated with a Finite Element (FE) model of the middle ear, in which similar alterations were simulated.
Results
The measured and calculated METF showed frequency and perforation size dependent losses at all perforation locations. Starting at low frequencies, the loss expanded to higher frequencies with increased perforation size. In direct comparison, posterior TM perforations affected the transmission properties to a larger degree than anterior perforations. The asymmetry of the TM causes the malleus-incus complex to rotate and results in larger deflections in the posterior TM quadrants than in the anterior TM quadrants. Simulations in the FE model with a sealed cavity show that small perforations lead to a decrease in TM rigidity and thus to an increase in oscillation amplitude of the TM mainly above 1 kHz.
Conclusion
Size and location of TM perforations have a characteristic influence on the METF. The correlation of the experimental LDV measurements with an FE model contributes to a better understanding of the pathologic mechanisms of middle-ear diseases. If small perforations with significant HL are observed in daily clinical practice, additional middle ear pathologies should be considered. Further investigations on the loss of TM pretension due to perforations may be informative.
The incudo-malleal joint (IMJ) in the human middle ear is a true diarthrodial joint and it has been known that the flexibility of this joint does not contribute to better middle-ear sound transmission. Previous studies have proposed that a gliding motion between the malleus and the incus at this joint prevents the transmission of large displacements of the malleus to the incus and stapes and thus contributes to the protection of the inner ear as an immediate response against large static pressure changes. However, dynamic behavior of this joint under static pressure changes has not been fully revealed. In this study, effects of the flexibility of the IMJ on middle-ear sound transmission under static pressure difference between the middle-ear cavity and the environment were investigated. Experiments were performed in human cadaveric temporal bones with static pressures in the range of +/- 2 kPa being applied to the ear canal (relative to middle-ear cavity). Vibrational motions of the umbo and the stapes footplate center in response to acoustic stimulation (0.2-8 kHz) were measured using a 3D-Laser Doppler vibrometer for (1) the natural IMJ and (2) the IMJ with experimentally-reduced flexibility. With the natural condition of the IMJ, vibrations of the umbo and the stapes footplate center under static pressure loads were attenuated at low frequencies below the middle-ear resonance frequency as observed in previous studies. After the flexibility of the IMJ was reduced, additional attenuations of vibrational motion were observed for the umbo under positive static pressures in the ear canal (EC) and the stapes footplate center under both positive and negative static EC pressures. The additional attenuation of vibration reached 4~7 dB for the umbo under positive static EC pressures and the stapes footplate center under negative EC pressures, and 7~11 dB for the stapes footplate center under positive EC pressures. The results of this study indicate an adaptive mechanism of the flexible IMJ in the human middle ear to changes of static EC pressure by reducing the attenuation of the middle-ear sound transmission. Such results are expected to be used for diagnosis of the IMJ stiffening and to be applied to design of middle-ear prostheses.
Knochenleitungshörgerät
(2021)
Bei einem Knochenleitungshörgerät (1) umfassend eine Audioeinheit (2) und einen Hautkontaktteil (3), wobei der Hautkontaktteil (3) eine Klebeschicht (4) zur Befestigung auf der Haut (5) eines Benutzers, ein erstes Kupplungsstück (6) zum Verbinden mit einem gegengleichen zweiten Kupplungsstück (7) der Audioeinheit (2), und einen Schwingungsübertragungsteil (8) mit einer Kontaktfläche (9) zum direkten Kontakt mit der Haut (5) aufweist, wird vorgeschlagen, dass das Schwingungsübertragungsteil (8) in einem Anpassungszustand durch Andrücken an die Haut (5) mit einer ersten Anpresskraft plastisch verformbar ist, dass das Schwingungsübertragungsteil (8) in einem Betriebszustand im Wesentlichen formstabil ist, und dass das Schwingungsübertragungsteil (8) durch einen Verfestigungsschritt von dem Anpassungszustand in den Betriebszustand verbringbar ist.