The focus of the developed maturity model was set on processes. The concept of the widespread CMM and its practices has been transferred to the perioperative domain and the concept of the new maturity model. Additional optimization goals and technological as well as networking-specific aspects enable a process- and object-focused view of the maturity model in order to ensure broad coverage of different subareas. The evaluation showed that the model is applicable to the perioperative field. Adjustments and extensions of the maturity model are future steps to improve the rating and classification of the new maturity model.
Access to clinical information during interventions is an important aspect to support the surgeon and his team in the OR. The OR-Pad research project aims at displaying clinically relevant information close to the patient during surgery. With the OR-Pad system, the surgeon shall be able to access case-specific information, displayed on a sterile-packaged, portable display device. Therefore, information shall be prepared before surgery and also be available afterwards. The project follows an user-centered design process. Within the third iteration, the interaction concept was finalized, resulting in an application that can be used in two modes, mobile and intraoperative, to support the surgeon before/after and during surgery, respectively. By supporting the surgeon perioperatively, it is expected to improve the information situation in the OR and thereby the quality of surgical results. Based on this concept, the system architecture was designed in detail, using a client-server architecture. Components, communication interfaces, exchanged data, and intended standards for data exchange of the OR-Pad system including connecting systems were conceived. Expert interviews by using a clickable prototype were conducted to evaluate the concepts.
Towards Automated Surgical Documentation using automatically generated checklists from BPMN models
(2021)
The documentation of surgeries is usually created from memory only after the operation, which is an additional effort for the surgeon and afflicted with the possibility of imprecisely, shortend reports. The display of process steps in the form of checklists and the automatic creation of surgical documentation from the completed process steps could serve as a reminder, standardize the surgical procedure and save time for the surgeon. Based on two works from Reutlingen University, which implemented the creation of dynamic checklists from Business Process Modelling Notation (BPMN) models and the storage of times at which a process step was completed, a prototype was developed for an android tablet, to expand the dynamic checklists by functions such as uploading photos and files, manual user entries, the interception of foreseeable deviations from the normal course of operations and the automatic creation of OR documentation.
Purpose: Medical processes can be modeled using different methods and notations.Currently used modeling systems like Business Process Model and Notation (BPMN) are not capable of describing the highly flexible and variable medical processes in sufficient detail.
Methods: We combined two modeling systems, Business Process Management (BPM) and Adaptive Case Management (ACM), to be able to model non-deterministic medical processes. We used the new Standards Case Management Model and Notation (CMMN) and Decision Management Notation (DMN).
Results: First, we explain how CMMN, DMN and BPMN could be used to model non-deterministic medical processes. We applied this methodology to model 79 cataract operations provided by University Hospital Leipzig, Germany, and four cataract operations provided by University Eye Hospital Tuebingen, Germany. Our model consists of 85 tasks and about 20 decisions in BPMN. We were able to expand the system with more complex situations that might appear during an intervention.
Conclusion: An effective modeling of the cataract intervention is possible using the combination of BPM and ACM. The combination gives the possibility to depict complex processes with complex decisions. This combination allows a significant advantage for modeling perioperative processes.
Physicians in interventional radiology are exposed to high physical stress. To avoid negative long-term effects resulting from unergonomic working conditions, we demonstrated the feasibility of a system that gives feedback about unergonomic
situations arising during the intervention based on the Azure Kinect camera. The overall feasibility of the approach could be shown.
Purpose
Context awareness in the operating room (OR) is important to realize targeted assistance to support actors during surgery. A situation recognition system (SRS) is used to interpret intraoperative events and derive an intraoperative situation from these. To achieve a modular system architecture, it is desirable to de-couple the SRS from other system components. This leads to the need of an interface between such an SRS and context-aware systems (CAS). This work aims to provide an open standardized interface to enable loose coupling of the SRS with varying CAS to allow vendor-independent device orchestrations.
Methods
A requirements analysis investigated limiting factors that currently prevent the integration of CAS in today's ORs. These elicited requirements enabled the selection of a suitable base architecture. We examined how to specify this architecture with the constraints of an interoperability standard. The resulting middleware was integrated into a prototypic SRS and our system for intraoperative support, the OR-Pad, as exemplary CAS for evaluating whether our solution can enable context-aware assistance during simulated orthopedical interventions.
Results
The emerging Service-oriented Device Connectivity (SDC) standard series was selected to specify and implement a middleware for providing the interpreted contextual information while the SRS and CAS are loosely coupled. The results were verified within a proof of concept study using the OR-Pad demonstration scenario. The fulfillment of the CAS’ requirements to act context-aware, conformity to the SDC standard series, and the effort for integrating the middleware in individual systems were evaluated. The semantically unambiguous encoding of contextual information depends on the further standardization process of the SDC nomenclature. The discussion of the validity of these results proved the applicability and transferability of the middleware.
Conclusion
The specified and implemented SDC-based middleware shows the feasibility of loose coupling an SRS with unknown CAS to realize context-aware assistance in the OR.
Ultra wideband real-time locating system for tracking people and devices in the operating room
(2022)
Position tracking within the OR could be one possible input for intraoperative situation recognition. Our approach demonstrates a Real-time Locating System (RTLS) using the Ultra Wideband (UWB) technology to determine the position of people or objects. The UWB RTLS was integrated into the research OR at Reutlingen University and the system’s settings were optimized regarding the four factors accuracy, susceptibility to interference, range, and latency. Therefore, different parameters were adapted and the effects on the factors were compared. Goodtracking quality could be achieved under optimal settings. These results indicate that a UWB RTLS is well suited to determine the position of people and devices in our setting. The feasibility of the system needsto be evaluated under real OR conditions.
With the progress of technology in modern hospitals, an intelligent perioperative situation recognition will gain more relevance due to its potential to substantially improve surgical workflows by providing situation knowledge in real-time. Such knowledge can be extracted from image data by machine learning techniques but poses a privacy threat to the staff’s and patients’ personal data. De-identification is a possible solution for removing visual sensitive information. In this work, we developed a YOLO v3 based prototype to detect sensitive areas in the image in real-time. These are then deidentified using common image obfuscation techniques. Our approach shows that it is principle suitable for de-identifying sensitive data in OR images and contributes to a privacyrespectful way of processing in the context of situation recognition in the OR.
One of the key challenges for automatic assistance is the support of actors in the operating room depending on the status of the procedure. Therefore, context information collected in the operating room is used to gain knowledge about the current situation. In literature, solutions already exist for specific use cases, but it is doubtful to what extent these approaches can be transferred to other conditions. We conducted a comprehensive literature research on existing situation recognition systems for the intraoperative area, covering 274 articles and 95 cross-references published between 2010 and 2019. We contrasted and compared 58 identified approaches based on defined aspects such as used sensor data or application area. In addition, we discussed applicability and transferability. Most of the papers focus on video data for recognizing situations within laparoscopic and cataract surgeries. Not all of the approaches can be used online for real-time recognition. Using different methods, good results with recognition accuracies above 90% could be achieved. Overall, transferability is less addressed. The applicability of approaches to other circumstances seems to be possible to a limited extent. Future research should place a stronger focus on adaptability. The literature review shows differences within existing approaches for situation recognition and outlines research trends. Applicability and transferability to other conditions are less addressed in current work.
What might the attendee be able to do after being in your session?
Our work shows how to connect intra-operative devices via IEEE 11073 Service-oriented Device Connectivity (SDC).
Description of the Problem or Gap
Standardized device communication is essential for interoperability, availability of device data, and therefore for the intelligent operating room (OR) and arising solutions. The SDC standard was developed to make information from medical devices available in a uniform manner and enable interoperability. Existing devices are rarely SDC-capable and need interfaces to be interoperable via SDC.
Methods: What did you do to address the problem or gap?
We conceived an SDC-based architecture consisting of a service provider and service consumer. In our concept, the service provider is connected to the medical device and capable to translate the proprietary protocol of the device into SDC and vice versa. The service consumer is used to request or send information via the SDC protocol to the service provider and can function as a uniform bidirectional interface (e.g. for displaying or controlling). This concept was exemplarily demonstrated with the patient monitor MX800 of Philips to retrieve the device data (e.g. vital parameters) via SDC and partly for the operating light marLED X of KLS Martin Group.
Results: What was the outcome(s) of what you did to address the problem or gap?
The patient monitor MX800 was connected to a Raspberry Pi (RPi) via LAN, on which the service provider is running. The python script on the RPi establishes a connection to the monitor and translates incoming and outgoing messages from the proprietary protocol to SDC and vice versa to/from the service consumer. The service consumer is running on a laptop and acts as a simulation for different kinds of systems that want to get vital parameters or other information from the patient monitor. The operating light marLED X was connected to an RPi via USB-to-RS232. A python script on the RPi establishes a connection to the light and makes it possible via proprietary commands to get information of the light (e.g. status) and to control it (e.g. toggle the light, increment the intensity). A translation to SDC is not integrated yet.
Discussion of Results
Our practical implementation shows that medical devices can be accessed via external connections to get device data and control the device via commands. The example SDC implementation of the patient monitor MX800 makes it possible to request its data via the standardized communication protocol SDC. This is also possible for the operating light marLED X if its proprietary protocol is analyzed to be translatable to/from SDC. This would allow to control the device from an external system, or automatically depending on the status of the ongoing procedure. The advantage is, that existing intra-operative devices can be extended by a service provider which is capable of translating the proprietary protocol of the device in SDC and vice versa. This enables interoperability and an intelligent OR that, for example, is aware of all devices, their status, and data and can use this information to optimally support the surgeons and their team (e.g. provision of information, automated documentation). This interoperability allows that future innovations merely need to understand the SDC protocol instead of all vendor-dependent communication protocols.
Conclusion
Standardized device communication is essential to reach interoperability, and therefore intelligent ORs. Our contribution addresses the possibility of subsequently making medical devices SDC-capable. This may eliminate the need of understanding all the different proprietary protocols when developing new innovative solutions for the OR.