This project aims to evaluate existing big data infrastructures for their applicability in the operating room to support medical staff with context-sensitive systems. Requirements for the system design were generated. The project compares different data mining technologies, interfaces, and software system infrastructures with a focus on their usefulness in the peri-operative setting. The lambda architecture was chosen for the proposed system design, which will provide data for both postoperative analysis and real-time support during surgery.
Introduction: Even if there is a standard procedure of CI surgery, especially in pediatric surgery surgical steps often differ individually due to anatomical variations, malformations or unforseen events. This is why every surgical report should be created individually, which takes time and relies on the correct memory of the surgeon. A standardized recording of intraoperative data and subsequent storage as well as text processing would therefore be desirable and provides the basis for subsequent data processing, e.g. in the context of research or quality assurance.
Method: In cooperation with Reutlingen University, we conducted a workflow analysis of the prototype of a semi-automatic checklist tool. Based on automatically generated checklists generated from BPMN models a prototype user interface was developed for an android tablet. Functions such as uploading photos and files, manual user entries, the interception of foreseeable deviations from the normal course of operations and the automatic creation of OP documentation could be implemented. The system was tested in a remote usability test on a petrous bone model.
Result: The user interface allows a simple intuitive handling, which can be well implemented in the intraoperative setting. Clinical data as well as surgical steps could be individually recorded and saved via DICOM. An automatic surgery report could be created and saved.
Summary: The use of a dynamic checklist tool facilitates the capture, storage and processing of surgical data. Further applications in clinical practice are pending.
Recent advances in artificial intelligence have enabled promising applications in neurosurgery that can enhance patient outcomes and minimize risks. This paper presents a novel system that utilizes AI to aid neurosurgeons in precisely identifying and localizing brain tumors. The system was trained on a dataset of brain MRI scans and utilized deep learning algorithms for segmentation and classification. Evaluation of the system on a separate set of brain MRI scans demonstrated an average Dice similarity coefficient of 0.87. The system was also evaluated through a user experience test involving the Department of Neurosurgery at the University Hospital Ulm, with results showing significant improvements in accuracy, efficiency, and reduced cognitive load and stress levels. Additionally, the system has demonstrated adaptability to various surgical scenarios and provides personalized guidance to users. These findings indicate the potential for AI to enhance the quality of neurosurgical interventions and improve patient outcomes. Future work will explore integrating this system with robotic surgical tools for minimally invasive surgeries.
Motivation
In order to enable context-aware behavior of surgical assistance systems, the acquisition of various information about the current intraoperative situation is crucial. To achieve this, the complex task of situation recognition can be delegated to a specialized system. Consequently, a standardized interface is required for the seamless transfer of the recognized contextual information to the assistance systems, enabling them to adapt accordingly.
Methods
Our group analyzed four medical interface standards to determine their suitability for exchanging intraoperative contextual information. The assessment was based on a harmonized data and service model derived from the requirements of expected context-aware use cases. The Digital Imaging and Communications in Medicine (DICOM) and IEEE 11073 for Service-oriented Device Connectivity (SDC) were identified as the most appropriate standards.
Results
We specified how DICOM Unified Procedure Steps (UPS), can be used to effectively communicate contextual information. We proposed the inclusion of attributes to formalize different granularity levels of the surgical workflow.
Conclusions
DICOM UPS SOP classes can be used for the exchange of intraoperative contextual information between a situation recognition system and surgical assistance systems. This can pave the way for vendor-independent context awareness in the OR, leading to targeted assistance of the surgical team and an improvement of the surgical workflow.
Purpose
For the modeling, execution, and control of complex, non-standardized intraoperative processes, a modeling language is needed that reflects the variability of interventions. As the established Business Process Model and Notation (BPMN) reaches its limits in terms of flexibility, the Case Management Model and Notation (CMMN) was considered as it addresses weakly structured processes.
Methods
To analyze the suitability of the modeling languages, BPMN and CMMN models of a Robot-Assisted Minimally Invasive Esophagectomy and Cochlea Implantation were derived and integrated into a situation recognition workflow. Test cases were used to contrast the differences and compare the advantages and disadvantages of the models concerning modeling, execution, and control. Furthermore, the impact on transferability was investigated.
Results
Compared to BPMN, CMMN allows flexibility for modeling intraoperative processes while remaining understandable. Although more effort and process knowledge are needed for execution and control within a situation recognition system, CMMN enables better transferability of the models and therefore the system. Concluding, CMMN should be chosen as a supplement to BPMN for flexible process parts that can only be covered insufficiently by BPMN, or otherwise as a replacement for the entire process.
Conclusion
CMMN offers the flexibility for variable, weakly structured process parts, and is thus suitable for surgical interventions. A combination of both notations could allow optimal use of their advantages and support the transferability of the situation recognition system.
Physicians in interventional radiology are exposed to high physical stress. To avoid negative long-term effects resulting from unergonomic working conditions, we demonstrated the feasibility of a system that gives feedback about unergonomic
situations arising during the intervention based on the Azure Kinect camera. The overall feasibility of the approach could be shown.
Ultra wideband real-time locating system for tracking people and devices in the operating room
(2022)
Position tracking within the OR could be one possible input for intraoperative situation recognition. Our approach demonstrates a Real-time Locating System (RTLS) using the Ultra Wideband (UWB) technology to determine the position of people or objects. The UWB RTLS was integrated into the research OR at Reutlingen University and the system’s settings were optimized regarding the four factors accuracy, susceptibility to interference, range, and latency. Therefore, different parameters were adapted and the effects on the factors were compared. Goodtracking quality could be achieved under optimal settings. These results indicate that a UWB RTLS is well suited to determine the position of people and devices in our setting. The feasibility of the system needsto be evaluated under real OR conditions.
With the progress of technology in modern hospitals, an intelligent perioperative situation recognition will gain more relevance due to its potential to substantially improve surgical workflows by providing situation knowledge in real-time. Such knowledge can be extracted from image data by machine learning techniques but poses a privacy threat to the staff’s and patients’ personal data. De-identification is a possible solution for removing visual sensitive information. In this work, we developed a YOLO v3 based prototype to detect sensitive areas in the image in real-time. These are then deidentified using common image obfuscation techniques. Our approach shows that it is principle suitable for de-identifying sensitive data in OR images and contributes to a privacyrespectful way of processing in the context of situation recognition in the OR.
What might the attendee be able to do after being in your session?
Our work shows how to connect intra-operative devices via IEEE 11073 Service-oriented Device Connectivity (SDC).
Description of the Problem or Gap
Standardized device communication is essential for interoperability, availability of device data, and therefore for the intelligent operating room (OR) and arising solutions. The SDC standard was developed to make information from medical devices available in a uniform manner and enable interoperability. Existing devices are rarely SDC-capable and need interfaces to be interoperable via SDC.
Methods: What did you do to address the problem or gap?
We conceived an SDC-based architecture consisting of a service provider and service consumer. In our concept, the service provider is connected to the medical device and capable to translate the proprietary protocol of the device into SDC and vice versa. The service consumer is used to request or send information via the SDC protocol to the service provider and can function as a uniform bidirectional interface (e.g. for displaying or controlling). This concept was exemplarily demonstrated with the patient monitor MX800 of Philips to retrieve the device data (e.g. vital parameters) via SDC and partly for the operating light marLED X of KLS Martin Group.
Results: What was the outcome(s) of what you did to address the problem or gap?
The patient monitor MX800 was connected to a Raspberry Pi (RPi) via LAN, on which the service provider is running. The python script on the RPi establishes a connection to the monitor and translates incoming and outgoing messages from the proprietary protocol to SDC and vice versa to/from the service consumer. The service consumer is running on a laptop and acts as a simulation for different kinds of systems that want to get vital parameters or other information from the patient monitor. The operating light marLED X was connected to an RPi via USB-to-RS232. A python script on the RPi establishes a connection to the light and makes it possible via proprietary commands to get information of the light (e.g. status) and to control it (e.g. toggle the light, increment the intensity). A translation to SDC is not integrated yet.
Discussion of Results
Our practical implementation shows that medical devices can be accessed via external connections to get device data and control the device via commands. The example SDC implementation of the patient monitor MX800 makes it possible to request its data via the standardized communication protocol SDC. This is also possible for the operating light marLED X if its proprietary protocol is analyzed to be translatable to/from SDC. This would allow to control the device from an external system, or automatically depending on the status of the ongoing procedure. The advantage is, that existing intra-operative devices can be extended by a service provider which is capable of translating the proprietary protocol of the device in SDC and vice versa. This enables interoperability and an intelligent OR that, for example, is aware of all devices, their status, and data and can use this information to optimally support the surgeons and their team (e.g. provision of information, automated documentation). This interoperability allows that future innovations merely need to understand the SDC protocol instead of all vendor-dependent communication protocols.
Conclusion
Standardized device communication is essential to reach interoperability, and therefore intelligent ORs. Our contribution addresses the possibility of subsequently making medical devices SDC-capable. This may eliminate the need of understanding all the different proprietary protocols when developing new innovative solutions for the OR.
One of the key challenges for automatic assistance is the support of actors in the operating room depending on the status of the procedure. Therefore, context information collected in the operating room is used to gain knowledge about the current situation. In literature, solutions already exist for specific use cases, but it is doubtful to what extent these approaches can be transferred to other conditions. We conducted a comprehensive literature research on existing situation recognition systems for the intraoperative area, covering 274 articles and 95 cross-references published between 2010 and 2019. We contrasted and compared 58 identified approaches based on defined aspects such as used sensor data or application area. In addition, we discussed applicability and transferability. Most of the papers focus on video data for recognizing situations within laparoscopic and cataract surgeries. Not all of the approaches can be used online for real-time recognition. Using different methods, good results with recognition accuracies above 90% could be achieved. Overall, transferability is less addressed. The applicability of approaches to other circumstances seems to be possible to a limited extent. Future research should place a stronger focus on adaptability. The literature review shows differences within existing approaches for situation recognition and outlines research trends. Applicability and transferability to other conditions are less addressed in current work.