A clinically useful system for individual continuous health data monitoring needs an architecture that takes into account all relevant medical and technical conditions. The requirements for a health app to support such a system are collected, and a vendor independent architecture is designed that allows the collection of vital data from arbitrary wearables using a smartphone. A prototypical implementation for the main scenario shows the feasibility of the approach.
Zur Unterstützung des Operateurs wird eine patientennahe Informationsanzeige entwickelt, die kontextrelevante Informationen entsprechend der aktuellen Situation bereitstellen kann. Hierfür soll eine Situationserkennung konzipiert werden, die auf unterschiedliche intraoperative Prozesse übertragen werden kann. Ziel der adaptiven Situationserkennung ist das Erkennen spezifischer Situationen durch intraoperative Informationen unterschiedlicher Datenquellen im Operationssaal. Innerhalb der Datenerhebung und -analyse wurden Anwendungsfälle für die Situationserkennung definiert sowie chirurgische Prozessmodelle erstellt, die intraoperative Ereignisse abbilden. Auf Basis dieser Informationen wurde ein Konzept entworfen, das sich zunächst auf die Erkennung abstrakter generalisierter Phasen, unabhängig vom Eingriff, fokussiert und sich Schritt für Schritt auf granulare Prozessschritte spezifizieren lässt. Diese Flexibilität soll die Übertragbarkeit des Konzepts auf intraoperative Prozesse ermöglichen und den Operateur dadurch gezielt mit kontextrelevanten Informationen unterstützen. Das Konzept wird in zukünftigen Schritten weiterentwickelt.
OR-Pad - Entwicklung eines Prototyps zur sterilen Informationsanzeige am OP-Situs : meeting abstract
(2019)
Hintergrund: Oftmals werden Informationen aus der Krankenakte oder von Bildgebungsverfahren nur auf recht weit vom Operationsgebiet entfernten Monitoren, außerhalb der ergonomischen Sichtachse des Operateurs, dargestellt. Dies führt dazu, dass relevante Informationen übersehen werden oder ihr Informationspotenzial nicht ausgeschöpft werden kann. In Papierform mitgenommene Notizen befinden sich während der OP außerhalb des sterilen Bereichs und sind dadurch für den Operateur nicht ohne Weiteres zugänglich. Auch bei intraoperativen Einträgen für die OP Dokumentation ist der Operateur auf die Mithilfe der Assistenz angewiesen. Durch die zusätzlichen Kommunikationswege entstehen dabei ein personeller und zeitlicher Mehraufwand und das Fehlerpotenzial nimmt zu. Das anwendungsorientierte Forschungsprojekt OR-Pad - Nutzung von portablen Informationsanzeigen im Operationssaal - soll dem Operateur zu einem verbesserten Informationsfluss verhelfen. Die Idee entstand aus der klinischen Routine der Anatomie und Urologie des Universitätsklinikums Tübingen und wird nun durch Fördermittel vom Ministerium für Wissenschaft, Forschung und Kunst Baden-Württemberg sowie vom Europäischen Fonds für regionale Entwicklung an der Hochschule Reutlingen zu einem High Fidelity-Prototypen weiterentwickelt.
Ziel: Ziel des OR-Pad Projekts ist es, während einer OP zum aktuellen Zeitpunkt klinisch relevante Informationen in unmittelbarer Nähe zum Operateur darzustellen. Mithilfe des Systems soll der Informationsfluss zwischen dem Eingriff sowie dessen Vor- und Nachbereitung optimiert werden. Der Operateur soll vorab relevante Informationen, wie aktuelle Röntgenbilder oder persönliche Notizen, zur intraoperativen Anzeige auswählen können, die dann am OP-Situs auf einer sterilen Informationsanzeige dargestellt werden. Durch die Positionierung soll eine ergonomische Sichtachse sowie die direkte Interaktion mit dem System ermöglicht werden. Kontextrelevante Informationen sollen basierend auf dem aktuellen OP-Verlauf durch die Entwicklung einer Situationserkennung automatisch bereitgestellt werden. Zur Optimierung des Informationsflusses gehört ebenfalls die Unterstützung der OP-Dokumentation. Für diese sollen während des Eingriffs manuell vom Operateur sowie automatisch vom System Einträge, wie Zeitpunkte oder intraoperative Aufnahmen, erstellt werden. Aus diesen soll nach dem Eingriff die OP-Dokumentation generiert und damit der Prozess qualitativer und zeiteffizienter gestaltet werden.
Methodik: Zur Erreichung des Ziels werden zunächst die klinischen Anforderungen spezifiziert und in ein Lastenheft überführt. Hierfür werden Interviews und Beobachtungen bei mehreren Interventionen durchgeführt. Nach dem User-Centered-Designprozess werden Personas und Nutzungsszenarien entworfen und mit klinischen Projektpartnern in mehreren Iterationen evaluiert. Es gilt eine Informationsarchitektur aufzubauen, die eine Einbettung klinischer Informationssysteme sowie Bild- und Gerätedaten aus dem OP-Netzwerk erlaubt. Eine Situationserkennung, basierend auf Prozessmodellen, soll zur Abschätzung des Operationsfortschritts entwickelt werden. Zur Befestigung der Informationsanzeige sollen geeignete Haltemechanismen eingesetzt werden. Das OR-Pad System soll laufend im Lehr- und Forschungs-OP der Hochschule Reutlingen getestet und im Sinne agiler Produktentwicklung mit den klinischen Projektpartnern abgestimmt werden. Der finale Funktionsprototyp soll abschließend in den Versuchs-OPs der Anatomie Tübingen getestet und evaluiert werden.
Ergebnisse: Über eine erste Datenerhebung mittels Contextual Inquiry konnten erste Anforderungen an das OR-Pad System erfasst werden, woraus ein Low-Fidelity-Prototyp resultierte. Die Evaluation über Experteninterviews führte in die zweite Iteration, in der das Konzept entsprechend der Ergebnisse angepasst wurde. Über Hospitationen am Uniklinikum Tübingen fand eine weitere Datenerhebung zur Erstellung von Szenarien für die intraoperativen Anwendungsfälle statt. Anhand der Anforderungen wurde ein Konzept für die Benutzerschnittstelle entworfen, die im weiteren Verlauf mit den klinischen Projektpartnern evaluiert wird.
One of the key challenges for automatic assistance is the support of actors in the operating room depending on the status of the procedure. Therefore, context information collected in the operating room is used to gain knowledge about the current situation. In literature, solutions already exist for specific use cases, but it is doubtful to what extent these approaches can be transferred to other conditions. We conducted a comprehensive literature research on existing situation recognition systems for the intraoperative area, covering 274 articles and 95 cross-references published between 2010 and 2019. We contrasted and compared 58 identified approaches based on defined aspects such as used sensor data or application area. In addition, we discussed applicability and transferability. Most of the papers focus on video data for recognizing situations within laparoscopic and cataract surgeries. Not all of the approaches can be used online for real-time recognition. Using different methods, good results with recognition accuracies above 90% could be achieved. Overall, transferability is less addressed. The applicability of approaches to other circumstances seems to be possible to a limited extent. Future research should place a stronger focus on adaptability. The literature review shows differences within existing approaches for situation recognition and outlines research trends. Applicability and transferability to other conditions are less addressed in current work.
Continuous monitoring of individual vital parameters can provide information for the assessment of one’s health and indications of medical problems in the context of personalized medicine. Correlations between parameters and health issues are to be evaluated. As one project in this topic area, a telemedicine platform is implemented to gather data of outpatients via wearables and accumulate them for physicians and researchers to review. This work extracts requirements, draws use case scenarios, and shows the current system architecture consisting of a patient application, a physician application with a web server, and a backend server application. In further work, the prototype will assist to develop a vendor-free and open monitoring solution. A conclusion on functionality and usability will be evaluated in an imminent first study.
Physicians in interventional radiology are exposed to high physical stress. To avoid negative long-term effects resulting from unergonomic working conditions, we demonstrated the feasibility of a system that gives feedback about unergonomic
situations arising during the intervention based on the Azure Kinect camera. The overall feasibility of the approach could be shown.
Purpose
Context awareness in the operating room (OR) is important to realize targeted assistance to support actors during surgery. A situation recognition system (SRS) is used to interpret intraoperative events and derive an intraoperative situation from these. To achieve a modular system architecture, it is desirable to de-couple the SRS from other system components. This leads to the need of an interface between such an SRS and context-aware systems (CAS). This work aims to provide an open standardized interface to enable loose coupling of the SRS with varying CAS to allow vendor-independent device orchestrations.
Methods
A requirements analysis investigated limiting factors that currently prevent the integration of CAS in today's ORs. These elicited requirements enabled the selection of a suitable base architecture. We examined how to specify this architecture with the constraints of an interoperability standard. The resulting middleware was integrated into a prototypic SRS and our system for intraoperative support, the OR-Pad, as exemplary CAS for evaluating whether our solution can enable context-aware assistance during simulated orthopedical interventions.
Results
The emerging Service-oriented Device Connectivity (SDC) standard series was selected to specify and implement a middleware for providing the interpreted contextual information while the SRS and CAS are loosely coupled. The results were verified within a proof of concept study using the OR-Pad demonstration scenario. The fulfillment of the CAS’ requirements to act context-aware, conformity to the SDC standard series, and the effort for integrating the middleware in individual systems were evaluated. The semantically unambiguous encoding of contextual information depends on the further standardization process of the SDC nomenclature. The discussion of the validity of these results proved the applicability and transferability of the middleware.
Conclusion
The specified and implemented SDC-based middleware shows the feasibility of loose coupling an SRS with unknown CAS to realize context-aware assistance in the OR.
Ultra wideband real-time locating system for tracking people and devices in the operating room
(2022)
Position tracking within the OR could be one possible input for intraoperative situation recognition. Our approach demonstrates a Real-time Locating System (RTLS) using the Ultra Wideband (UWB) technology to determine the position of people or objects. The UWB RTLS was integrated into the research OR at Reutlingen University and the system’s settings were optimized regarding the four factors accuracy, susceptibility to interference, range, and latency. Therefore, different parameters were adapted and the effects on the factors were compared. Goodtracking quality could be achieved under optimal settings. These results indicate that a UWB RTLS is well suited to determine the position of people and devices in our setting. The feasibility of the system needsto be evaluated under real OR conditions.
With the progress of technology in modern hospitals, an intelligent perioperative situation recognition will gain more relevance due to its potential to substantially improve surgical workflows by providing situation knowledge in real-time. Such knowledge can be extracted from image data by machine learning techniques but poses a privacy threat to the staff’s and patients’ personal data. De-identification is a possible solution for removing visual sensitive information. In this work, we developed a YOLO v3 based prototype to detect sensitive areas in the image in real-time. These are then deidentified using common image obfuscation techniques. Our approach shows that it is principle suitable for de-identifying sensitive data in OR images and contributes to a privacyrespectful way of processing in the context of situation recognition in the OR.
What might the attendee be able to do after being in your session?
Our work shows how to connect intra-operative devices via IEEE 11073 Service-oriented Device Connectivity (SDC).
Description of the Problem or Gap
Standardized device communication is essential for interoperability, availability of device data, and therefore for the intelligent operating room (OR) and arising solutions. The SDC standard was developed to make information from medical devices available in a uniform manner and enable interoperability. Existing devices are rarely SDC-capable and need interfaces to be interoperable via SDC.
Methods: What did you do to address the problem or gap?
We conceived an SDC-based architecture consisting of a service provider and service consumer. In our concept, the service provider is connected to the medical device and capable to translate the proprietary protocol of the device into SDC and vice versa. The service consumer is used to request or send information via the SDC protocol to the service provider and can function as a uniform bidirectional interface (e.g. for displaying or controlling). This concept was exemplarily demonstrated with the patient monitor MX800 of Philips to retrieve the device data (e.g. vital parameters) via SDC and partly for the operating light marLED X of KLS Martin Group.
Results: What was the outcome(s) of what you did to address the problem or gap?
The patient monitor MX800 was connected to a Raspberry Pi (RPi) via LAN, on which the service provider is running. The python script on the RPi establishes a connection to the monitor and translates incoming and outgoing messages from the proprietary protocol to SDC and vice versa to/from the service consumer. The service consumer is running on a laptop and acts as a simulation for different kinds of systems that want to get vital parameters or other information from the patient monitor. The operating light marLED X was connected to an RPi via USB-to-RS232. A python script on the RPi establishes a connection to the light and makes it possible via proprietary commands to get information of the light (e.g. status) and to control it (e.g. toggle the light, increment the intensity). A translation to SDC is not integrated yet.
Discussion of Results
Our practical implementation shows that medical devices can be accessed via external connections to get device data and control the device via commands. The example SDC implementation of the patient monitor MX800 makes it possible to request its data via the standardized communication protocol SDC. This is also possible for the operating light marLED X if its proprietary protocol is analyzed to be translatable to/from SDC. This would allow to control the device from an external system, or automatically depending on the status of the ongoing procedure. The advantage is, that existing intra-operative devices can be extended by a service provider which is capable of translating the proprietary protocol of the device in SDC and vice versa. This enables interoperability and an intelligent OR that, for example, is aware of all devices, their status, and data and can use this information to optimally support the surgeons and their team (e.g. provision of information, automated documentation). This interoperability allows that future innovations merely need to understand the SDC protocol instead of all vendor-dependent communication protocols.
Conclusion
Standardized device communication is essential to reach interoperability, and therefore intelligent ORs. Our contribution addresses the possibility of subsequently making medical devices SDC-capable. This may eliminate the need of understanding all the different proprietary protocols when developing new innovative solutions for the OR.