Refine
Document Type
- Journal article (6)
- Conference proceeding (5)
- Doctoral Thesis (1)
Has full text
- yes (12) (remove)
Is part of the Bibliography
- yes (12)
Institute
- Technik (12)
Publisher
Wave-like differential equations occur in many engineering applications. Here the engineering setup is embedded into the framework of functional analysis of modern mathematical physics. After an overview, the –Hilbert space approach to free Euler–Bernoulli bending vibrations of a beam in one spatial dimension is investigated. We analyze in detail the corresponding positive, selfadjoint differential operators of 4-th order associated to the boundary conditions in statics. A comparison with free string wave swinging is outlined.
Perforations of the tympanic membrane (TM) can occur as a result of injury or inflammation of the middle ear. These perforations can lead to conductive hearing loss (HL), where in some cases the magnitude of HL exceeds that attributable to the observed TM perforation alone. We aim with this study to better understand the effects of location and size of TM perforations on the sound transmitting properties of the middle ear.
The middle ear transfer function (METF) of six human temporal bones (TB; freshly frozen specimen of body donors) were compared before and after perforation of the TM at different locations (anterior or posterior lower quadrant) and of different sizes (1mm, ¼ of the TM, ½ of the TM, and full ablation). The
METF were correlated with a Finite Element (FE) model of the middle ear, in which similar alterations were simulated.
The measured and simulated FE model METFs exhibited frequency and perforation size dependent amplitude losses at all locations and severities. In direct comparison, posterior TM perforations affected the transmission properties to a larger degree than perforations of the anterior quadrant. This could possibly be caused by an asymmetry of the TM, where the malleus-incus complex rotates and results in larger deflections in the posterior TM half than in the anterior TM half. The FE model of the TM with a sealed cavity suggest that small perforations result in a decrease of TM rigidity and thus to an increase in oscillation amplitude of the TM, mostly above 1 kHz.
The location and size of TM perforations influence the METF in a reproducible way. Correlating our data with the FE model could help to better understand the pathologic mechanisms of middle-ear diseases. If small TM perforations with uncharacteristically significant HL are observed in daily clinical practice, additional middle ear pathologies should be considered. Further investigations on the loss of TM pretension due to perforations may be informative.
The hearing contact lens® (HCL) is a new type of hearing aid devices. One of its main components is a piezo-electric actuator. In order to evaluate and maximize the HCL's performance, a model of the HCL coupled to a middle-ear model was developed using finite element approach. The model was validated step by step starting with the HCL only. To validate the HCL model, vibrational measurements on the HCL were performed using a laser-doppler-vibrometer (LDV). Then, a silicone cap was placed onto the HCL to provide an interface between the HCL and the tympanic membrane of the middle-ear model, and additional LDV measurements on temporal bones were performed to validate the coupled model that was used to evaluate the equivalent sound pressure of the HCL. Moreover, a de-eper insight was gained into the contact between the HCL and tympanic membrane and its effects on the HCL performance. The model can be used to investigate the sensitivity of geometrical and material parameters with respect to performance measures of the HCL and evaluate the feedback behavior.
Hearing contact lens (HCL) is a new type of hearing aid devices. One of its main components is a piezo-electric actuator (PEA). In order to evaluate and maximizethe HCL´s performance, a model of the HCL coupled to the middle ear was developed using finite element (FE)approach. To validate the model, vibrational measurements on the HCL and temporal bones were performed using a Laser-Doppler-Vibrometer (LDV). The model was validated step by step starting with HCL only. Then a silicone cap was fitted onto the HCL to provide an interface between the HCL and the tympanic membrane. The HCL was placed on the tympanic membrane and additional measurements were performed to validate the coupled model. The model was used to evaluate the sensitivity of geometrical and material parameters with respect to performance measures of the HCL. Moreover, deeper insight was gained into the feedback behavior, which causes whistling sounds, and the contact between the HCL and tympanic membrane.
Purpose
Injury or inflammation of the middle ear often results in the persistent tympanic membrane (TM) perforations, leading to conductive hearing loss (HL). However, in some cases the magnitude of HL exceeds that attributable by the TM perforation alone. The aim of the study is to better understand the effects of location and size of TM perforations on the sound transmission properties of the middle ear.
Methods
The middle ear transfer functions (METF) of six human temporal bones (TB) were compared before and after perforating the TM at different locations (anterior or posterior lower quadrant) and to different degrees (1 mm, ¼ of the TM, ½ of the TM, and full ablation). The sound-induced velocity of the stapes footplate was measured using single-point laser-Doppler-vibrometry (LDV). The METF were correlated with a Finite Element (FE) model of the middle ear, in which similar alterations were simulated.
Results
The measured and calculated METF showed frequency and perforation size dependent losses at all perforation locations. Starting at low frequencies, the loss expanded to higher frequencies with increased perforation size. In direct comparison, posterior TM perforations affected the transmission properties to a larger degree than anterior perforations. The asymmetry of the TM causes the malleus-incus complex to rotate and results in larger deflections in the posterior TM quadrants than in the anterior TM quadrants. Simulations in the FE model with a sealed cavity show that small perforations lead to a decrease in TM rigidity and thus to an increase in oscillation amplitude of the TM mainly above 1 kHz.
Conclusion
Size and location of TM perforations have a characteristic influence on the METF. The correlation of the experimental LDV measurements with an FE model contributes to a better understanding of the pathologic mechanisms of middle-ear diseases. If small perforations with significant HL are observed in daily clinical practice, additional middle ear pathologies should be considered. Further investigations on the loss of TM pretension due to perforations may be informative.
The incudo-malleal joint (IMJ) in the human middle ear is a true diarthrodial joint and it has been known that the flexibility of this joint does not contribute to better middle-ear sound transmission. Previous studies have proposed that a gliding motion between the malleus and the incus at this joint prevents the transmission of large displacements of the malleus to the incus and stapes and thus contributes to the protection of the inner ear as an immediate response against large static pressure changes. However, dynamic behavior of this joint under static pressure changes has not been fully revealed. In this study, effects of the flexibility of the IMJ on middle-ear sound transmission under static pressure difference between the middle-ear cavity and the environment were investigated. Experiments were performed in human cadaveric temporal bones with static pressures in the range of +/- 2 kPa being applied to the ear canal (relative to middle-ear cavity). Vibrational motions of the umbo and the stapes footplate center in response to acoustic stimulation (0.2-8 kHz) were measured using a 3D-Laser Doppler vibrometer for (1) the natural IMJ and (2) the IMJ with experimentally-reduced flexibility. With the natural condition of the IMJ, vibrations of the umbo and the stapes footplate center under static pressure loads were attenuated at low frequencies below the middle-ear resonance frequency as observed in previous studies. After the flexibility of the IMJ was reduced, additional attenuations of vibrational motion were observed for the umbo under positive static pressures in the ear canal (EC) and the stapes footplate center under both positive and negative static EC pressures. The additional attenuation of vibration reached 4~7 dB for the umbo under positive static EC pressures and the stapes footplate center under negative EC pressures, and 7~11 dB for the stapes footplate center under positive EC pressures. The results of this study indicate an adaptive mechanism of the flexible IMJ in the human middle ear to changes of static EC pressure by reducing the attenuation of the middle-ear sound transmission. Such results are expected to be used for diagnosis of the IMJ stiffening and to be applied to design of middle-ear prostheses.
Investigation of tympanic membrane influences on middle-ear impedance measurements and simulations
(2020)
This study simulates acoustic impedance measurements in the human ear canal and investigates error influences due to improperly accounted evanescence in the probe’s near field, cross-section area changes, curvature of the ear canal, and pressure inhomogeneities across the tympanic membrane, which arise mainly at frequencies above 10 kHz. Evanescence results from strongly damped modes of higher order, which can only be found in the near field of the sound source and are excited due to sharp cross-sectional changes as they occur at the transition from the probe loudspeaker to the ear canal. This means that different impedances are measured depending on the probe design. The influence of evanescence cannot be eliminated completely from measurements, however, it can be reduced by a probe design with larger distance between speaker and microphone. A completely different approach to account for the influence of evanescence is to evaluate impedance measurements with the help of a finite element model, which takes the precise arrangement of microphone and speaker in the measurement into account. The latter is shown in this study exemplary on impedance measurements at a tube terminated with a steel plate. Furthermore, the influences of shape changes of the tympanic membrane and ear canal curvature on impedance are investigated.
This study describes a non-contact measuring and system identification procedure for evaluating inhomogeneous stiffness and damping characteristics of the annular ligament in the physiological amplitude and frequency range without the application of large static external forces that can cause unnatural displacements of the stapes. To verify the procedure, measurements were first conducted on a steel beam. Then, measurements on an individual human cadaveric temporal bone sample were performed. The estimated results support the inhomogeneous stiffness and damping distribution of the annular ligament and are in a good agreement with the multiphoton microscopy results which show that the posterior-inferior corner of the stapes footplate is the stiffest region of the annular ligament.
This study describes a non-contact measuring and parameter identification procedure designed to evaluate inhomogeneous stiffness and damping characteristics of the annular ligament in the physiological amplitude and frequency range without the application of large static external forces that can cause unnatural displacements of the stapes. To verify the procedure, measurements were first conducted on a steel beam. Then, measurements on an individual human cadaveric temporal bone sample were performed. The estimated results support the inhomogeneous stiffness and damping distribution of the annular ligament and are in a good agreement with the multiphoton microscopy results which show that the posterior-inferior corner of the stapes footplate is the stiffest region of the annular ligament. This method can potentially help to establish a correlation between stiffness and damping characteristics of the annular ligament and inertia properties of the stapes and, thus, help to reduce the number of independent parameters in the model-based hearing diagnosis.
Due to the large interindividual variances and the poor optical accessibility of the ear, the specificity of hearing diagnostics today is severely restricted to a certain clinical picture and quantitative assessment. Often only a yes or no decision is possible, which depends strongly on the subjective assessment of the ENT physician. A novel approach, in which objectively obtainable, non invasive audiometric measurements are evaluated using a numerical middle ear model, makes it possible to make the hidden middle ear properties visible and quantifiable. The central topic of this paper is a novel parameter identification algorithm that combines inverse fuzzy arithmetic with an artificial neural network in order to achieve a coherent diagnostic overall picture in the comparison of model and measurement. Its usage is shown at a pathological pattern called malleus fixation where the upper ligament of the malleus is pathologically stiffened.