Stent graft visualization and planning tool for endovascular surgery using finite element analysis
(2014)
Purpose: A new approach to optimize stent graft selection for endovascular aortic repair is the use of finite element analysis. Once the finite element model is created and solved, a software module is needed to view the simulation results in the clinical work environment. A new tool for Interpretation of simulation results, named Medical Postprocessor, that enables comparison of different stent graft configurations and products was designed, implemented and tested. Methods Aortic endovascular stent graft ring forces and sealing states in the vessel landing zone of three different configurations were provided in a surgical planning software using the Medical Imaging Interaction Tool Kit (MITK) Software system. For data interpretation, software modules for 2D and 3D presentations were implemented. Ten surgeons evaluated the software features of the Medical Postprocessor. These surgeons performed usability tests and answered questionnaires based on their experience with the system.
Results: The Medical Postprocessor visualization system enabled vascular surgeons to determine the configuration with the highest overall fixation force in 16 ± 6 s, best proximal sealing in 56±24 s and highest proximal fixation force in 38 ± 12 s. The majority considered the multiformat data provided helpful and found the Medical Postprocessor to be an efficient decision support system for stent graft selection. The evaluation of the user interface results in an ISONORMconform user interface (113.5 points).
Conclusion: The Medical Postprocessor visualization Software tool for analyzing stent graft properties was evaluated by vascular surgeons. The results show that the software can assist the interpretation of simulation results to optimize stent graft configuration and sizing.
Purpose
Supporting the surgeon during surgery is one of the main goals of intelligent ORs. The OR-Pad project aims to optimize the information flow within the perioperative area. A shared information space should enable appropriate preparation and provision of relevant information at any time before, during, and after surgery.
Methods
Based on previous work on an interaction concept and system architecture for the sterile OR-Pad system, we designed a user interface for mobile and intraoperative (stationary) use, focusing on the most important functionalities like clear information provision to reduce information overload. The concepts were transferred into a high-fidelity prototype for demonstration purposes. The prototype was evaluated from different perspectives, including a usability study.
Results
The prototype’s central element is a timeline displaying all available case information chronologically, like radiological images, labor findings, or notes. This information space can be adapted for individual purposes (e.g., highlighting a tumor, filtering for own material). With the mobile and intraoperative mode of the system, relevant information can be added, preselected, viewed, and extended during the perioperative process. Overall, the evaluation showed good results and confirmed the vision of the information system.
Conclusion
The high-fidelity prototype of the information system OR-Pad focuses on supporting the surgeon via a timeline making all available case information accessible before, during, and after surgery. The information space can be personalized to enable targeted support. Further development is reasonable to optimize the approach and address missing or insufficient aspects, like the holding arm and sterility concept or new desired features.
Checklists are a valuable tool to ensure process quality and quality of care. To ensure proper integration in clinical processes, it would be desirable to generate checklists directly from formal process descriptions. Those checklists could also be used for user interaction in context-aware surgical assist systems. We built a tool to automatically convert Business Process Model and Notation (BPMN) process models to checklists displayed as HTML websites. Gateways representing decisions are mapped to checklist items that trigger dynamic content loading based on the placed checkmark. The usability of the resulting system was positively evaluated regarding comprehensibility and end-user friendliness.
Hintergrund: Endoskopische Operationsverfahren haben sich als Goldstandard in der Nasennebenhöhlen-(NNH-)Chirurgie etabliert. Den sich daraus ergebenden Herausforderungen für die chirurgische Ausbildung kann durch den Einsatz von Virtuelle-Realität-(VR-)Trainingssimulatoren begegnet werden. Bislang wurde eine Reihe von Simulatoren für NNH-Operationen entwickelt. Frühere Studien im Hinblick auf den Trainingseffekt wurden jedoch nur mit medizinisch vorgebildeten Probanden durchgeführt oder es wurde nicht über dessen zeitlichen Verlauf berichtet.
Methoden: Ein NNH-CT-Datensatz wurde nach der Segmentierung in ein 3-dimensionales, polygonales Oberflächenmodell überführt und mithilfe von originalem Fotomaterial texturiert. Die Interaktion mit der virtuellen Umgebung erfolgte über ein haptisches Eingabegerät. Während der Simulation wurden die Parameter Eingriffsdauer und Fehleranzahl erfasst. Zehn Probanden absolvierten jeweils eine Trainingseinheit bestehend aus je 5 Übungsdurchläufen an 10 aufeinanderfolgenden Tagen.
Ergebnisse: Vier Probanden verringerten die benötigte Zeit um mehr als 60% im Verlauf des Übungszeitraums. Vier der Probanden verringerten ihre Fehleranzahl um mehr als 60%. Acht von 10 Probanden zeigten eine Verbesserung bezüglich beider Parameter. Im Median wurde im gesamten gemessenen Zeitraum die Dauer des Eingriffs um 46 Sekunden und die Fehleranzahl um 191 reduziert. Die Überprüfung eines Zusammenhangs zwischen den 2 Parametern ergab eine positive Korrelation.
Schlussfolgerung: Zusammenfassend lässt sich feststellen, dass das Training am NNH-Simulator auch bei unerfahrenen Personen die Performance beträchtlich verbessert, sowohl in Bezug auf die Dauer als auch auf die Genauigkeit des Eingriffs.
In this paper a method for the generation of gSPM with ontology-based generalization was presented. The resulting gSPM was modeled with BPMN/BPMNsix in an efficient way and could be executed with BPMN workflow engines. In the next step the implementation of resource concepts, anatomical structures, and transition probabilities for workflow execution will be realized.
Ultra wideband real-time locating system for tracking people and devices in the operating room
(2022)
Position tracking within the OR could be one possible input for intraoperative situation recognition. Our approach demonstrates a Real-time Locating System (RTLS) using the Ultra Wideband (UWB) technology to determine the position of people or objects. The UWB RTLS was integrated into the research OR at Reutlingen University and the system’s settings were optimized regarding the four factors accuracy, susceptibility to interference, range, and latency. Therefore, different parameters were adapted and the effects on the factors were compared. Goodtracking quality could be achieved under optimal settings. These results indicate that a UWB RTLS is well suited to determine the position of people and devices in our setting. The feasibility of the system needsto be evaluated under real OR conditions.
Background and purpose: Transapical aortic valve replacement (TAVR) is a recent minimally invasive surgical treatment technique for elderly and high-risk patients with severe aortic stenosis. In this paper,a simple and accurate image-based method is introduced to aid the intra-operative guidance of TAVR procedure under 2-D X-ray fluoroscopy.
Methods: The proposed method fuses a 3-D aortic mesh model and anatomical valve landmarks with live 2-D fluoroscopic images. The 3-D aortic mesh model and landmarks are reconstructed from interventional X-ray C-arm CT system, and a target area for valve implantation is automatically estimated using these aortic mesh models.Based on template-based tracking approach, the overlay of visualized 3-D aortic mesh model, land-marks and target area of implantation is updated onto fluoroscopic images by approximating the aortic root motion from a pigtail catheter motion without contrast agent. Also, a rigid intensity-based registration algorithm is used to track continuously the aortic root motion in the presence of contrast agent.Furthermore, a sensorless tracking of the aortic valve prosthesis is provided to guide the physician to perform the appropriate placement of prosthesis into the estimated target area of implantation.
Results: Retrospective experiments were carried out on fifteen patient datasets from the clinical routine of the TAVR. The maximum displacement errors were less than 2.0 mm for both the dynamic overlay of aortic mesh models and image-based tracking of the prosthesis, and within the clinically accepted ranges. Moreover, high success rates of the proposed method were obtained above 91.0% for all tested patient datasets.
Conclusion: The results showed that the proposed method for computer-aided TAVR is potentially a helpful tool for physicians by automatically defining the accurate placement position of the prosthesis during the surgical procedure.
Intra-operative fluoroscopy-guided assistance system for transcatheter aortic valve implantation
(2014)
A new surgical assistance system has been developed to assist the correct positioning of the AVP during transapical TAVI. The developed assistance system automatically defines the target area for implanting the AVP under live 2-D fluoroscopy guidance. Moreover, this surgical assistance system works with low levels of contrast agent for the final deployment of AVP, reducing therefore long-term negative effects, such as renal failure in the elderly and high-risk patients.
Towards Automated Surgical Documentation using automatically generated checklists from BPMN models
(2021)
The documentation of surgeries is usually created from memory only after the operation, which is an additional effort for the surgeon and afflicted with the possibility of imprecisely, shortend reports. The display of process steps in the form of checklists and the automatic creation of surgical documentation from the completed process steps could serve as a reminder, standardize the surgical procedure and save time for the surgeon. Based on two works from Reutlingen University, which implemented the creation of dynamic checklists from Business Process Modelling Notation (BPMN) models and the storage of times at which a process step was completed, a prototype was developed for an android tablet, to expand the dynamic checklists by functions such as uploading photos and files, manual user entries, the interception of foreseeable deviations from the normal course of operations and the automatic creation of OR documentation.
Purpose
For the modeling, execution, and control of complex, non-standardized intraoperative processes, a modeling language is needed that reflects the variability of interventions. As the established Business Process Model and Notation (BPMN) reaches its limits in terms of flexibility, the Case Management Model and Notation (CMMN) was considered as it addresses weakly structured processes.
Methods
To analyze the suitability of the modeling languages, BPMN and CMMN models of a Robot-Assisted Minimally Invasive Esophagectomy and Cochlea Implantation were derived and integrated into a situation recognition workflow. Test cases were used to contrast the differences and compare the advantages and disadvantages of the models concerning modeling, execution, and control. Furthermore, the impact on transferability was investigated.
Results
Compared to BPMN, CMMN allows flexibility for modeling intraoperative processes while remaining understandable. Although more effort and process knowledge are needed for execution and control within a situation recognition system, CMMN enables better transferability of the models and therefore the system. Concluding, CMMN should be chosen as a supplement to BPMN for flexible process parts that can only be covered insufficiently by BPMN, or otherwise as a replacement for the entire process.
Conclusion
CMMN offers the flexibility for variable, weakly structured process parts, and is thus suitable for surgical interventions. A combination of both notations could allow optimal use of their advantages and support the transferability of the situation recognition system.