Context awareness in the operating room (OR) is important to realize targeted assistance to support actors during surgery. A situation recognition system (SRS) is used to interpret intraoperative events and derive an intraoperative situation from these. To achieve a modular system architecture, it is desirable to de-couple the SRS from other system components. This leads to the need of an interface between such an SRS and context-aware systems (CAS). This work aims to provide an open standardized interface to enable loose coupling of the SRS with varying CAS to allow vendor-independent device orchestrations.
Methods
A requirements analysis investigated limiting factors that currently prevent the integration of CAS in today's ORs. These elicited requirements enabled the selection of a suitable base architecture. We examined how to specify this architecture with the constraints of an interoperability standard. The resulting middleware was integrated into a prototypic SRS and our system for intraoperative support, the OR-Pad, as exemplary CAS for evaluating whether our solution can enable context-aware assistance during simulated orthopedical interventions.
Results
The emerging Service-oriented Device Connectivity (SDC) standard series was selected to specify and implement a middleware for providing the interpreted contextual information while the SRS and CAS are loosely coupled. The results were verified within a proof of concept study using the OR-Pad demonstration scenario. The fulfillment of the CAS’ requirements to act context-aware, conformity to the SDC standard series, and the effort for integrating the middleware in individual systems were evaluated. The semantically unambiguous encoding of contextual information depends on the further standardization process of the SDC nomenclature. The discussion of the validity of these results proved the applicability and transferability of the middleware.
Conclusion
The specified and implemented SDC-based middleware shows the feasibility of loose coupling an SRS with unknown CAS to realize context-aware assistance in the OR.
The paper describes how eye-tracking can be used to explore electronic patient records (EPR) in a sterile environment. As an information display, we used a system that we developed for the presentation of patient data and for supporting surgical hand disinfection. The eye-tracking was performed using the Tobii Eye Tracker 4C, and the connection between the eye-tracker and the HTML website was realized using the Tobii EyeX Chrome Extension. Interactions with the EPR are triggered by fixations of icons. The interaction was working as intended, but test persons reported a high mental load while using the system.
Ultra wideband real-time locating system for tracking people and devices in the operating room
(2022)
Position tracking within the OR could be one possible input for intraoperative situation recognition. Our approach demonstrates a Real-time Locating System (RTLS) using the Ultra Wideband (UWB) technology to determine the position of people or objects. The UWB RTLS was integrated into the research OR at Reutlingen University and the system’s settings were optimized regarding the four factors accuracy, susceptibility to interference, range, and latency. Therefore, different parameters were adapted and the effects on the factors were compared. Goodtracking quality could be achieved under optimal settings. These results indicate that a UWB RTLS is well suited to determine the position of people and devices in our setting. The feasibility of the system needsto be evaluated under real OR conditions.
With the progress of technology in modern hospitals, an intelligent perioperative situation recognition will gain more relevance due to its potential to substantially improve surgical workflows by providing situation knowledge in real-time. Such knowledge can be extracted from image data by machine learning techniques but poses a privacy threat to the staff’s and patients’ personal data. De-identification is a possible solution for removing visual sensitive information. In this work, we developed a YOLO v3 based prototype to detect sensitive areas in the image in real-time. These are then deidentified using common image obfuscation techniques. Our approach shows that it is principle suitable for de-identifying sensitive data in OR images and contributes to a privacyrespectful way of processing in the context of situation recognition in the OR.
Intraoperative imaging can assist neurosurgeons to define brain tumours and other surrounding brain structures. Interventional ultrasound (iUS) is a convenient modality with fast scan times. However, iUS data may suffer from noise and artefacts which limit their interpretation during brain surgery. In this work, we use two deep learning networks, namely UNet and TransUNet, to make automatic and accurate segmentation of the brain tumour in iUS data. Experiments were conducted on a dataset of 27 iUS volumes. The outcomes show that using a transformer with UNet is advantageous providing an efficient segmentation modelling long-range dependencies between each iUS image. In particular, the enhanced TransUNet was able to predict cavity segmentation in iUS data with an inference rate of more than 125 FPS. These promising results suggest that deep learning networks can be successfully deployed to assist neurosurgeons in the operating room.
Glioblastomas are the most aggressive fast-growing primary brain cancer which originate in the glial cells of the brain. Accurate identification of the malignant brain tumor and its sub-regions is still one of the most challenging problems in medical image segmentation. The Brain Tumor Segmentation Challenge (BraTS) has been a popular benchmark for automatic brain glioblastomas segmentation algorithms since its initiation. In this year, BraTS 2021 challenge provides the largest multi-parametric (mpMRI) dataset of 2,000 pre-operative patients. In this paper, we propose a new aggregation of two deep learning frameworksnamely, DeepSeg and nnU-Net for automatic glioblastoma recognition in pre-operative mpMRI. Our ensemble method obtains Dice similarity scores of 92.00, 87.33, and 84.10 and Hausdorff Distances of 3.81, 8.91, and 16.02 for the enhancing tumor, tumor core, and whole tumor regions, respectively, on the BraTS 2021 validation set, ranking us among the top ten teams. These experimental findings provide evidence that it can be readily applied clinically and thereby aiding in the brain cancer prognosis, therapy planning, and therapy response monitoring. A docker image for reproducing our segmentation results is available online at (https://hub.docker.com/r/razeineldin/deepseg21).
What might the attendee be able to do after being in your session?
Our work shows how to connect intra-operative devices via IEEE 11073 Service-oriented Device Connectivity (SDC).
Description of the Problem or Gap
Standardized device communication is essential for interoperability, availability of device data, and therefore for the intelligent operating room (OR) and arising solutions. The SDC standard was developed to make information from medical devices available in a uniform manner and enable interoperability. Existing devices are rarely SDC-capable and need interfaces to be interoperable via SDC.
Methods: What did you do to address the problem or gap?
We conceived an SDC-based architecture consisting of a service provider and service consumer. In our concept, the service provider is connected to the medical device and capable to translate the proprietary protocol of the device into SDC and vice versa. The service consumer is used to request or send information via the SDC protocol to the service provider and can function as a uniform bidirectional interface (e.g. for displaying or controlling). This concept was exemplarily demonstrated with the patient monitor MX800 of Philips to retrieve the device data (e.g. vital parameters) via SDC and partly for the operating light marLED X of KLS Martin Group.
Results: What was the outcome(s) of what you did to address the problem or gap?
The patient monitor MX800 was connected to a Raspberry Pi (RPi) via LAN, on which the service provider is running. The python script on the RPi establishes a connection to the monitor and translates incoming and outgoing messages from the proprietary protocol to SDC and vice versa to/from the service consumer. The service consumer is running on a laptop and acts as a simulation for different kinds of systems that want to get vital parameters or other information from the patient monitor. The operating light marLED X was connected to an RPi via USB-to-RS232. A python script on the RPi establishes a connection to the light and makes it possible via proprietary commands to get information of the light (e.g. status) and to control it (e.g. toggle the light, increment the intensity). A translation to SDC is not integrated yet.
Discussion of Results
Our practical implementation shows that medical devices can be accessed via external connections to get device data and control the device via commands. The example SDC implementation of the patient monitor MX800 makes it possible to request its data via the standardized communication protocol SDC. This is also possible for the operating light marLED X if its proprietary protocol is analyzed to be translatable to/from SDC. This would allow to control the device from an external system, or automatically depending on the status of the ongoing procedure. The advantage is, that existing intra-operative devices can be extended by a service provider which is capable of translating the proprietary protocol of the device in SDC and vice versa. This enables interoperability and an intelligent OR that, for example, is aware of all devices, their status, and data and can use this information to optimally support the surgeons and their team (e.g. provision of information, automated documentation). This interoperability allows that future innovations merely need to understand the SDC protocol instead of all vendor-dependent communication protocols.
Conclusion
Standardized device communication is essential to reach interoperability, and therefore intelligent ORs. Our contribution addresses the possibility of subsequently making medical devices SDC-capable. This may eliminate the need of understanding all the different proprietary protocols when developing new innovative solutions for the OR.
Background
Personalized medicine requires the integration and analysis of vast amounts of patient data to realize individualized care. With Surgomics, we aim to facilitate personalized therapy recommendations in surgery by integration of intraoperative surgical data and their analysis with machine learning methods to leverage the potential of this data in analogy to Radiomics and Genomics.
Methods
We defined Surgomics as the entirety of surgomic features that are process characteristics of a surgical procedure automatically derived from multimodal intraoperative data to quantify processes in the operating room. In a multidisciplinary team we discussed potential data sources like endoscopic videos, vital sign monitoring, medical devices and instruments and respective surgomic features. Subsequently, an online questionnaire was sent to experts from surgery and (computer) science at multiple centers for rating the features’ clinical relevance and technical feasibility.
Results
In total, 52 surgomic features were identified and assigned to eight feature categories. Based on the expert survey (n = 66 participants) the feature category with the highest clinical relevance as rated by surgeons was “surgical skill and quality of performance” for morbidity and mortality (9.0 ± 1.3 on a numerical rating scale from 1 to 10) as well as for long-term (oncological) outcome (8.2 ± 1.8). The feature category with the highest feasibility to be automatically extracted as rated by (computer) scientists was “Instrument” (8.5 ± 1.7). Among the surgomic features ranked as most relevant in their respective category were “intraoperative adverse events”, “action performed with instruments”, “vital sign monitoring”, and “difficulty of surgery”.
Conclusion
Surgomics is a promising concept for the analysis of intraoperative data. Surgomics may be used together with preoperative features from clinical data and Radiomics to predict postoperative morbidity, mortality and long-term outcome, as well as to provide tailored feedback for surgeons.
Hintergrund: Endoskopische Operationsverfahren haben sich als Goldstandard in der Nasennebenhöhlen-(NNH-)Chirurgie etabliert. Den sich daraus ergebenden Herausforderungen für die chirurgische Ausbildung kann durch den Einsatz von Virtuelle-Realität-(VR-)Trainingssimulatoren begegnet werden. Bislang wurde eine Reihe von Simulatoren für NNH-Operationen entwickelt. Frühere Studien im Hinblick auf den Trainingseffekt wurden jedoch nur mit medizinisch vorgebildeten Probanden durchgeführt oder es wurde nicht über dessen zeitlichen Verlauf berichtet.
Methoden: Ein NNH-CT-Datensatz wurde nach der Segmentierung in ein 3-dimensionales, polygonales Oberflächenmodell überführt und mithilfe von originalem Fotomaterial texturiert. Die Interaktion mit der virtuellen Umgebung erfolgte über ein haptisches Eingabegerät. Während der Simulation wurden die Parameter Eingriffsdauer und Fehleranzahl erfasst. Zehn Probanden absolvierten jeweils eine Trainingseinheit bestehend aus je 5 Übungsdurchläufen an 10 aufeinanderfolgenden Tagen.
Ergebnisse: Vier Probanden verringerten die benötigte Zeit um mehr als 60% im Verlauf des Übungszeitraums. Vier der Probanden verringerten ihre Fehleranzahl um mehr als 60%. Acht von 10 Probanden zeigten eine Verbesserung bezüglich beider Parameter. Im Median wurde im gesamten gemessenen Zeitraum die Dauer des Eingriffs um 46 Sekunden und die Fehleranzahl um 191 reduziert. Die Überprüfung eines Zusammenhangs zwischen den 2 Parametern ergab eine positive Korrelation.
Schlussfolgerung: Zusammenfassend lässt sich feststellen, dass das Training am NNH-Simulator auch bei unerfahrenen Personen die Performance beträchtlich verbessert, sowohl in Bezug auf die Dauer als auch auf die Genauigkeit des Eingriffs.
Purpose
Supporting the surgeon during surgery is one of the main goals of intelligent ORs. The OR-Pad project aims to optimize the information flow within the perioperative area. A shared information space should enable appropriate preparation and provision of relevant information at any time before, during, and after surgery.
Methods
Based on previous work on an interaction concept and system architecture for the sterile OR-Pad system, we designed a user interface for mobile and intraoperative (stationary) use, focusing on the most important functionalities like clear information provision to reduce information overload. The concepts were transferred into a high-fidelity prototype for demonstration purposes. The prototype was evaluated from different perspectives, including a usability study.
Results
The prototype’s central element is a timeline displaying all available case information chronologically, like radiological images, labor findings, or notes. This information space can be adapted for individual purposes (e.g., highlighting a tumor, filtering for own material). With the mobile and intraoperative mode of the system, relevant information can be added, preselected, viewed, and extended during the perioperative process. Overall, the evaluation showed good results and confirmed the vision of the information system.
Conclusion
The high-fidelity prototype of the information system OR-Pad focuses on supporting the surgeon via a timeline making all available case information accessible before, during, and after surgery. The information space can be personalized to enable targeted support. Further development is reasonable to optimize the approach and address missing or insufficient aspects, like the holding arm and sterility concept or new desired features.
Purpose
Artificial intelligence (AI), in particular deep learning (DL), has achieved remarkable results for medical image analysis in several applications. Yet the lack of human-like explanations of such systems is considered the principal restriction before utilizing these methods in clinical practice (Yang, Ye, & Xia, 2022).
Methods
Explainable Artificial Intelligence (XAI) provides a human-explainable and interpretable description of the “black-box” nature of DL (Gulum, Trombley, & Kantardzic, 2021). An effective XAI diagnosis generator, namely NeuroXAI (refer to Fig. 1), has been developed to extract 3D explanations from convolutional neural networks (CNN) models of brain gliomas (Zeineldin et al., 2022). By providing visual justification maps, NeuroXAI can help make DL models transparent and thus increase the trust of medical experts.
Results
NeuroXAI has been applied to two applications of the most widely investigated problems in brain imaging analysis, i.e. image classification and segmentation using magnetic resonance imaging (MRI). Visual attention maps of multiple XAI methods have been generated and compared for both applications, which could help to provide transparency about the performance of DL systems.
Conclusion
NeuroXAI helps to understand the prediction process of 3D CNN networks for brain glioma using human-understandable explanations. Results revealed that the investigated DL models behave in a logical human-like manner and can improve the analytical process of the MRI images systematically. Due to its open architecture, ease of implementation, and scalability to new XAI methods, NeuroXAI could be utilized to assist medical professionals in the detection and diagnosis of brain tumors. NeuroXAI code is publicly accessible at https://github.com/razeineldin/NeuroXAI
Access to clinical information during interventions is an important aspect to support the surgeon and his team in the OR. The OR-Pad research project aims at displaying clinically relevant information close to the patient during surgery. With the OR-Pad system, the surgeon shall be able to access case-specific information, displayed on a sterile-packaged, portable display device. Therefore, information shall be prepared before surgery and also be available afterwards. The project follows an user-centered design process. Within the third iteration, the interaction concept was finalized, resulting in an application that can be used in two modes, mobile and intraoperative, to support the surgeon before/after and during surgery, respectively. By supporting the surgeon perioperatively, it is expected to improve the information situation in the OR and thereby the quality of surgical results. Based on this concept, the system architecture was designed in detail, using a client-server architecture. Components, communication interfaces, exchanged data, and intended standards for data exchange of the OR-Pad system including connecting systems were conceived. Expert interviews by using a clickable prototype were conducted to evaluate the concepts.
Towards Automated Surgical Documentation using automatically generated checklists from BPMN models
(2021)
The documentation of surgeries is usually created from memory only after the operation, which is an additional effort for the surgeon and afflicted with the possibility of imprecisely, shortend reports. The display of process steps in the form of checklists and the automatic creation of surgical documentation from the completed process steps could serve as a reminder, standardize the surgical procedure and save time for the surgeon. Based on two works from Reutlingen University, which implemented the creation of dynamic checklists from Business Process Modelling Notation (BPMN) models and the storage of times at which a process step was completed, a prototype was developed for an android tablet, to expand the dynamic checklists by functions such as uploading photos and files, manual user entries, the interception of foreseeable deviations from the normal course of operations and the automatic creation of OR documentation.
Purpose
Computerized medical imaging processing assists neurosurgeons to localize tumours precisely. It plays a key role in recent image-guided neurosurgery. Hence, we developed a new open-source toolkit, namely Slicer-DeepSeg, for efficient and automatic brain tumour segmentation based on deep learning methodologies for aiding clinical brain research.
Methods
Our developed toolkit consists of three main components. First, Slicer-DeepSeg extends the 3D Slicer application and thus provides support for multiple data input/ output data formats and 3D visualization libraries. Second, Slicer core modules offer powerful image processing and analysis utilities. Third, the Slicer-DeepSeg extension provides a customized GUI for brain tumour segmentation using deep learning-based methods.
Results
The developed Slicer-DeepSeg was validated using a public dataset of high-grade glioma patients. The results showed that our proposed platform’s performance considerably outperforms other 3D Slicer cloud-based approaches.
Conclusions
Developed Slicer-DeepSeg allows the development of novel AI-assisted medical applications in neurosurgery. Moreover, it can enhance the outcomes of computer-aided diagnosis of brain tumours. Open-source Slicer-DeepSeg is available at github.com/razeineldin/Slicer-DeepSeg.
Die Bereitstellung klinischer Informationen im Operationssaal ist ein wichtiger Aspekt zur Unterstützung des chirurgischen Teams. Die roboter-assistierte Ösophagusresektion ist ein besonders komplexer Eingriff, der Potenzial zur workflowbasierten Unterstützung bietet. Wir präsentieren erste Ergebnisse der Entwicklung eines Checklisten-Tools mit der zugrundeliegenden Modellierung des chirurgischen Workflows und Informationsbedarf der Chirurgen. Das Checklisten-Tool zeigt hierfür die durchzuführenden Schritte chronologisch an und stellt zusätzliche Informationen kontextadaptiert bereit. Eine automatische Dokumentation von Start- und Endzeiten einzelner OP-Phasen und Schritte soll zukünftige Prozessanalysen der Operation ermöglichen.
A hybrid deep registration of MR scans to interventional ultrasound for neurosurgical guidance
(2021)
Despite the recent advances in image-guided neurosurgery, reliable and accurate estimation of the brain shift still remains one of the key challenges. In this paper, we propose an automated multimodal deformable registration method using hybrid learning-based and classical approaches to improve neurosurgical procedures. Initially, the moving and fixed images are aligned using classical affine transformation (MINC toolkit), and then the result is provided to the convolutional neural network, which predicts the deformation field using backpropagation. Subsequently, the moving image is transformed using the resultant deformation into a moved image. Our model was evaluated on two publicly available datasets: the retrospective evaluation of cerebral tumors (RESECT) and brain images of tumors for evaluation (BITE). The mean target registration errors have been reduced from 5.35 ± 4.29 to 0.99 ± 0.22 mm in the RESECT and from 4.18 ± 1.91 to 1.68 ± 0.65 mm in the BITE. Experimental results showed that our method improved the state-of-the-art in terms of both accuracy and runtime speed (170 ms on average). Hence, the proposed method provides a fast runtime for 3D MRI to intra-operative US pair in a GPU-based implementation, which shows a promise for its applicability in assisting the neurosurgical procedures compensating for brain shift.
Accurate and safe neurosurgical intervention can be affected by intra-operative tissue deformation, known as brain-shift. In this study, we propose an automatic, fast, and accurate deformable method, called iRegNet, for registering pre-operative magnetic resonance images to intra-operative ultrasound volumes to compensate for brain-shift. iRegNet is a robust end-to-end deep learning approach for the non-linear registration of MRI-iUS images in the context of image-guided neurosurgery. Pre-operative MRI (as moving image) and iUS (as fixed image) are first appended to our convolutional neural network, after which a non-rigid transformation field is estimated. The MRI image is then transformed using the output displacement field to the iUS coordinate system. Extensive experiments have been conducted on two multi-location databases, which are the BITE and the RESECT. Quantitatively, iRegNet reduced the mean landmark errors from pre-registration value of (4.18 ± 1.84 and 5.35 ± 4.19 mm) to the lowest value of (1.47 ± 0.61 and 0.84 ± 0.16 mm) for the BITE and RESECT datasets, respectively. Additional qualitative validation of this study was conducted by two expert neurosurgeons through overlaying MRI-iUS pairs before and after the deformable registration. Experimental findings show that our proposed iRegNet is fast and achieves state-of-the-art accuracies outperforming state-of-the-art approaches. Furthermore, the proposed iRegNet can deliver competitive results, even in the case of non-trained images as proof of its generality and can therefore be valuable in intra-operative neurosurgical guidance.
Context-aware systems to support actors in the operating room depending on the status of the intervention require knowledge about the current situation in the intra-operative area. In literature, solutions to achieve situation awareness already exist for specific use cases, but applicability and transferability to other conditions are less addressed. It is assumed that a unified solution that can be adapted to different processes and sensors would allow for greater flexibility, applicability, and thus transferability to different applications. To enable a flexible and intervention-independent system, this work proposes a concept for an adaptable situation recognition system. The system consists of four layers with several modular components for different functionalities. The feasibility is demonstrated via prototypical implementation and functional evaluation of a first basic framework prototype. Further development goal is the stepwise extension of the prototype.
Uncontrolled movements of laparoscopic instruments can lead to inadvertent injury of adjacent structures. The risk becomes evident when the dissecting instrument is located outside the field of view of the laparoscopic camera. Technical solutions to ensure patient safety are appreciated. The present work evaluated the feasibility of an automated binary classification of laparoscopic image data using Convolutional Neural Networks (CNN) to determine whether the dissecting instrument is located within the laparoscopic image section. A unique record of images was generated from six laparoscopic cholecystectomies in a surgical training environment to configure and train The CNN. By using a temporary version of the neural network, the annotation of the training image files could be automated and accelerated. A combination of oversampling and selective data augmentation was used to enlarge the fully labelled image data set and prevent loss of accuracy due to imbalanced class volumes. Subsequently the same approach was applied to the comprehensive, fully annotated Cholec80 database. The described process led to the generation of extensive and balanced training image data sets. The performance of the CNN-based binary classifiers was evaluated on separate test records from both databases. On our recorded data, an accuracy of 0.88 with regard to the safety-relevant classification was achieved. The subsequent evaluation on the Cholec80 data set yielded an accuracy of 0.84. The presented results demonstrate the feasibility of a binary classification of laparoscopic image data for the detection of adverse events in a surgical training environment using a specifically configured CNN architecture.
Purpose: Gliomas are the most common and aggressive type of brain tumors due to their infiltrative nature and rapid progression. The process of distinguishing tumor boundaries from healthy cells is still a challenging task in the clinical routine. Fluid attenuated inversion recovery (FLAIR) MRI modality can provide the physician with information about tumor infiltration. Therefore, this paper proposes a new generic deep learning architecture, namely DeepSeg, for fully automated detection and segmentation of the brain lesion using FLAIR MRI data.
Methods: The developed DeepSeg is a modular decoupling framework. It consists of two connected core parts based on an encoding and decoding relationship. The encoder part is a convolutional neural network (CNN) responsible for spatial information extraction. The resulting semantic map is inserted into the decoder part to get the full-resolution probability map. Based on modified U-Net architecture, different CNN models such as residual neural network (ResNet), dense convolutional network (DenseNet), and NASNet have been utilized in this study.
Results: The proposed deep learning architectures have been successfully tested and evaluated on-line based on MRI datasets of brain tumor segmentation (BraTS 2019) challenge, including s336 cases as training data and 125 cases for validation data. The dice and Hausdorff distance scores of obtained segmentation results are about 0.81 to 0.84 and 9.8 to 19.7 correspondingly.
Conclusion: This study showed successful feasibility and comparative performance of applying different deep learning models in a new DeepSeg framework for automated brain tumor segmentation in FLAIR MR images. The proposed DeepSeg is open source and freely available at https://github.com/razeineldin/DeepSeg/.