Refine
Year of publication
- 2019 (4) (remove)
Document Type
- Conference proceeding (3)
- Journal article (1)
Language
- English (4)
Has full text
- yes (4)
Is part of the Bibliography
- yes (4)
Institute
- Technik (4)
Publisher
Model-based hearing diagnosis based on wideband tympanometry measurements utilizing fuzzy arithmetic
(2019)
Today's audiometric methods for the diagnosis of middle ear disease are often based on a comparison of measurements with standard curves that represent the statistical range of normal hearing responses. Because of large inter-individual variances in the middle ear, especially in wideband tympanometry (WBT), specificity and quantitative evaluation are greatly restricted. A new model-based approach could transform today's predominantly qualitative hearing diganostics into a quantitative and tailored, patient-specific diagnosis, by evaluating WBT measurements with the aid of a middle-ear model. For this particular investigation, a finite element model of a human ear was used. It consisted of an acoustic ear canal and a tympanic cavity model, a middle-ear with detailed nonlinear models of the tympanic membrane and annular ligament, and a simplified inner-ear model. This model has made it possible to identify pathologies from measurements, by analyzing the parameters through senstivity studies and parameter clustering. Uncertainties due to the lack of knowledge, subjectivity in numerical implementation and model simplification are taken into account by the application of fuzzy arithmetic. The most confident parameter set can be determined by applying an inverse fuzzy method on the measurement data. The principle and the benefits of this model-based approach are illustrated by the example of a two-mass oscillator, and also by the simulation of the energy absorbance of an ear with malleus fixation, where the parameter changes that are introduced can be determined quantitatively through the system identification.
Due to the large interindividual variances and the poor optical accessibility of the ear, the specificity of hearing diagnostics today is severely restricted to a certain clinical picture and quantitative assessment. Often only a yes or no decision is possible, which depends strongly on the subjective assessment of the ENT physician. A novel approach, in which objectively obtainable, non invasive audiometric measurements are evaluated using a numerical middle ear model, makes it possible to make the hidden middle ear properties visible and quantifiable. The central topic of this paper is a novel parameter identification algorithm that combines inverse fuzzy arithmetic with an artificial neural network in order to achieve a coherent diagnostic overall picture in the comparison of model and measurement. Its usage is shown at a pathological pattern called malleus fixation where the upper ligament of the malleus is pathologically stiffened.
This study describes a non-contact measuring and parameter identification procedure designed to evaluate inhomogeneous stiffness and damping characteristics of the annular ligament in the physiological amplitude and frequency range without the application of large static external forces that can cause unnatural displacements of the stapes. To verify the procedure, measurements were first conducted on a steel beam. Then, measurements on an individual human cadaveric temporal bone sample were performed. The estimated results support the inhomogeneous stiffness and damping distribution of the annular ligament and are in a good agreement with the multiphoton microscopy results which show that the posterior-inferior corner of the stapes footplate is the stiffest region of the annular ligament. This method can potentially help to establish a correlation between stiffness and damping characteristics of the annular ligament and inertia properties of the stapes and, thus, help to reduce the number of independent parameters in the model-based hearing diagnosis.
This study describes a non-contact measuring and system identification procedure for evaluating inhomogeneous stiffness and damping characteristics of the annular ligament in the physiological amplitude and frequency range without the application of large static external forces that can cause unnatural displacements of the stapes. To verify the procedure, measurements were first conducted on a steel beam. Then, measurements on an individual human cadaveric temporal bone sample were performed. The estimated results support the inhomogeneous stiffness and damping distribution of the annular ligament and are in a good agreement with the multiphoton microscopy results which show that the posterior-inferior corner of the stapes footplate is the stiffest region of the annular ligament.