Introduction: Even if there is a standard procedure of CI surgery, especially in pediatric surgery surgical steps often differ individually due to anatomical variations, malformations or unforseen events. This is why every surgical report should be created individually, which takes time and relies on the correct memory of the surgeon. A standardized recording of intraoperative data and subsequent storage as well as text processing would therefore be desirable and provides the basis for subsequent data processing, e.g. in the context of research or quality assurance.
Method: In cooperation with Reutlingen University, we conducted a workflow analysis of the prototype of a semi-automatic checklist tool. Based on automatically generated checklists generated from BPMN models a prototype user interface was developed for an android tablet. Functions such as uploading photos and files, manual user entries, the interception of foreseeable deviations from the normal course of operations and the automatic creation of OP documentation could be implemented. The system was tested in a remote usability test on a petrous bone model.
Result: The user interface allows a simple intuitive handling, which can be well implemented in the intraoperative setting. Clinical data as well as surgical steps could be individually recorded and saved via DICOM. An automatic surgery report could be created and saved.
Summary: The use of a dynamic checklist tool facilitates the capture, storage and processing of surgical data. Further applications in clinical practice are pending.
Ultra wideband real-time locating system for tracking people and devices in the operating room
(2022)
Position tracking within the OR could be one possible input for intraoperative situation recognition. Our approach demonstrates a Real-time Locating System (RTLS) using the Ultra Wideband (UWB) technology to determine the position of people or objects. The UWB RTLS was integrated into the research OR at Reutlingen University and the system’s settings were optimized regarding the four factors accuracy, susceptibility to interference, range, and latency. Therefore, different parameters were adapted and the effects on the factors were compared. Goodtracking quality could be achieved under optimal settings. These results indicate that a UWB RTLS is well suited to determine the position of people and devices in our setting. The feasibility of the system needsto be evaluated under real OR conditions.
With the progress of technology in modern hospitals, an intelligent perioperative situation recognition will gain more relevance due to its potential to substantially improve surgical workflows by providing situation knowledge in real-time. Such knowledge can be extracted from image data by machine learning techniques but poses a privacy threat to the staff’s and patients’ personal data. De-identification is a possible solution for removing visual sensitive information. In this work, we developed a YOLO v3 based prototype to detect sensitive areas in the image in real-time. These are then deidentified using common image obfuscation techniques. Our approach shows that it is principle suitable for de-identifying sensitive data in OR images and contributes to a privacyrespectful way of processing in the context of situation recognition in the OR.
Towards Automated Surgical Documentation using automatically generated checklists from BPMN models
(2021)
The documentation of surgeries is usually created from memory only after the operation, which is an additional effort for the surgeon and afflicted with the possibility of imprecisely, shortend reports. The display of process steps in the form of checklists and the automatic creation of surgical documentation from the completed process steps could serve as a reminder, standardize the surgical procedure and save time for the surgeon. Based on two works from Reutlingen University, which implemented the creation of dynamic checklists from Business Process Modelling Notation (BPMN) models and the storage of times at which a process step was completed, a prototype was developed for an android tablet, to expand the dynamic checklists by functions such as uploading photos and files, manual user entries, the interception of foreseeable deviations from the normal course of operations and the automatic creation of OR documentation.
Intraoperative brain deformation, so called brain shift, affects the applicability of preoperative magnetic resonance imaging (MRI) data to assist the procedures of intraoperative ultrasound (iUS) guidance during neurosurgery. This paper proposes a deep learning-based approach for fast and accurate deformable registration of preoperative MRI to iUS images to correct brain shift. Based on the architecture of 3D convolutional neural networks, the proposed deep MRI-iUS registration method has been successfully tested and evaluated on the retrospective evaluation of cerebral tumors (RESECT) dataset. This study showed that our proposed method outperforms other registration methods in previous studies with an average mean squared error (MSE) of 85. Moreover, this method can register three 3D MRI-US pair in less than a second, improving the expected outcomes of brain surgery.
Purpose
For the modeling, execution, and control of complex, non-standardized intraoperative processes, a modeling language is needed that reflects the variability of interventions. As the established Business Process Model and Notation (BPMN) reaches its limits in terms of flexibility, the Case Management Model and Notation (CMMN) was considered as it addresses weakly structured processes.
Methods
To analyze the suitability of the modeling languages, BPMN and CMMN models of a Robot-Assisted Minimally Invasive Esophagectomy and Cochlea Implantation were derived and integrated into a situation recognition workflow. Test cases were used to contrast the differences and compare the advantages and disadvantages of the models concerning modeling, execution, and control. Furthermore, the impact on transferability was investigated.
Results
Compared to BPMN, CMMN allows flexibility for modeling intraoperative processes while remaining understandable. Although more effort and process knowledge are needed for execution and control within a situation recognition system, CMMN enables better transferability of the models and therefore the system. Concluding, CMMN should be chosen as a supplement to BPMN for flexible process parts that can only be covered insufficiently by BPMN, or otherwise as a replacement for the entire process.
Conclusion
CMMN offers the flexibility for variable, weakly structured process parts, and is thus suitable for surgical interventions. A combination of both notations could allow optimal use of their advantages and support the transferability of the situation recognition system.
There are several intra-operative use cases which require the surgeon to interact with medical devices. We used the Leap Motion Controller as input device and implemented two use-cases: 2D-Interaction (e.g. advancing EPR data) and selection of a value (e.g. room illumination brightness). The gesture detection was successful and we mapped its output to several devices and systems.
Background
Personalized medicine requires the integration and analysis of vast amounts of patient data to realize individualized care. With Surgomics, we aim to facilitate personalized therapy recommendations in surgery by integration of intraoperative surgical data and their analysis with machine learning methods to leverage the potential of this data in analogy to Radiomics and Genomics.
Methods
We defined Surgomics as the entirety of surgomic features that are process characteristics of a surgical procedure automatically derived from multimodal intraoperative data to quantify processes in the operating room. In a multidisciplinary team we discussed potential data sources like endoscopic videos, vital sign monitoring, medical devices and instruments and respective surgomic features. Subsequently, an online questionnaire was sent to experts from surgery and (computer) science at multiple centers for rating the features’ clinical relevance and technical feasibility.
Results
In total, 52 surgomic features were identified and assigned to eight feature categories. Based on the expert survey (n = 66 participants) the feature category with the highest clinical relevance as rated by surgeons was “surgical skill and quality of performance” for morbidity and mortality (9.0 ± 1.3 on a numerical rating scale from 1 to 10) as well as for long-term (oncological) outcome (8.2 ± 1.8). The feature category with the highest feasibility to be automatically extracted as rated by (computer) scientists was “Instrument” (8.5 ± 1.7). Among the surgomic features ranked as most relevant in their respective category were “intraoperative adverse events”, “action performed with instruments”, “vital sign monitoring”, and “difficulty of surgery”.
Conclusion
Surgomics is a promising concept for the analysis of intraoperative data. Surgomics may be used together with preoperative features from clinical data and Radiomics to predict postoperative morbidity, mortality and long-term outcome, as well as to provide tailored feedback for surgeons.
Stent graft visualization and planning tool for endovascular surgery using finite element analysis
(2014)
Purpose: A new approach to optimize stent graft selection for endovascular aortic repair is the use of finite element analysis. Once the finite element model is created and solved, a software module is needed to view the simulation results in the clinical work environment. A new tool for Interpretation of simulation results, named Medical Postprocessor, that enables comparison of different stent graft configurations and products was designed, implemented and tested. Methods Aortic endovascular stent graft ring forces and sealing states in the vessel landing zone of three different configurations were provided in a surgical planning software using the Medical Imaging Interaction Tool Kit (MITK) Software system. For data interpretation, software modules for 2D and 3D presentations were implemented. Ten surgeons evaluated the software features of the Medical Postprocessor. These surgeons performed usability tests and answered questionnaires based on their experience with the system.
Results: The Medical Postprocessor visualization system enabled vascular surgeons to determine the configuration with the highest overall fixation force in 16 ± 6 s, best proximal sealing in 56±24 s and highest proximal fixation force in 38 ± 12 s. The majority considered the multiformat data provided helpful and found the Medical Postprocessor to be an efficient decision support system for stent graft selection. The evaluation of the user interface results in an ISONORMconform user interface (113.5 points).
Conclusion: The Medical Postprocessor visualization Software tool for analyzing stent graft properties was evaluated by vascular surgeons. The results show that the software can assist the interpretation of simulation results to optimize stent graft configuration and sizing.