Publisher
Model-guided Therapy and Surgical Workflow Systems are two interrelated research fields, which have been developed separately in the last years. To make full use of both technologies, it is necessary to integrate them and connect them to Hospital Information Systems. We propose a framework for integration of Model-guided Therapy in Hospital Information Systems based on the Electronic Medical Record, and a taskbased Workflow Management System, which is suitable for clinical end users. Two prototypes - one based on Business Process Modeling Language, one based on the serum-board - are presented. From the experience with these prototypes, we developed a novel personalized visualization system for Surgical Workflows and Model-guided Therapy. Key challenges for further development are automated situation detection and a common communication infrastructure.
Intra-operative fluoroscopy-guided assistance system for transcatheter aortic valve implantation
(2014)
A new surgical assistance system has been developed to assist the correct positioning of the AVP during transapical TAVI. The developed assistance system automatically defines the target area for implanting the AVP under live 2-D fluoroscopy guidance. Moreover, this surgical assistance system works with low levels of contrast agent for the final deployment of AVP, reducing therefore long-term negative effects, such as renal failure in the elderly and high-risk patients.
There are several intra-operative use cases which require the surgeon to interact with medical devices. We used the Leap Motion Controller as input device and implemented two use-cases: 2D-Interaction (e.g. advancing EPR data) and selection of a value (e.g. room illumination brightness). The gesture detection was successful and we mapped its output to several devices and systems.
Stent graft visualization and planning tool for endovascular surgery using finite element analysis
(2014)
Purpose: A new approach to optimize stent graft selection for endovascular aortic repair is the use of finite element analysis. Once the finite element model is created and solved, a software module is needed to view the simulation results in the clinical work environment. A new tool for Interpretation of simulation results, named Medical Postprocessor, that enables comparison of different stent graft configurations and products was designed, implemented and tested. Methods Aortic endovascular stent graft ring forces and sealing states in the vessel landing zone of three different configurations were provided in a surgical planning software using the Medical Imaging Interaction Tool Kit (MITK) Software system. For data interpretation, software modules for 2D and 3D presentations were implemented. Ten surgeons evaluated the software features of the Medical Postprocessor. These surgeons performed usability tests and answered questionnaires based on their experience with the system.
Results: The Medical Postprocessor visualization system enabled vascular surgeons to determine the configuration with the highest overall fixation force in 16 ± 6 s, best proximal sealing in 56±24 s and highest proximal fixation force in 38 ± 12 s. The majority considered the multiformat data provided helpful and found the Medical Postprocessor to be an efficient decision support system for stent graft selection. The evaluation of the user interface results in an ISONORMconform user interface (113.5 points).
Conclusion: The Medical Postprocessor visualization Software tool for analyzing stent graft properties was evaluated by vascular surgeons. The results show that the software can assist the interpretation of simulation results to optimize stent graft configuration and sizing.
An operation room is a stressful work environment. Nevertheless, all involved persons have to work safely as there is no space for making mistakes. To ensure a high level of concentration and seamless interaction, all involved persons have to know their own tasks and tasks of their colleagues. The entire team must work synchronously at all times. However, the operation room (OR) is a noisy environment and the actors have to set their focus on their work. To optimize the overall workflow, a task manager supporting the team was developed. Each actor is equipped with a client terminal showing a summary of their own tasks. Moreover, a big screen displays all tasks of all actors. The architecture is a distributed system based on a communication framework that supports the interaction of all clients with the task manager. A prototype of the task manager and several clients have been developed and implemented. The system represents a proof-of-concept for further development. This paper describes the concept of the task manager.