Perforations of the tympanic membrane (TM) can occur as a result of injury or inflammation of the middle ear. These perforations can lead to conductive hearing loss (HL), where in some cases the magnitude of HL exceeds that attributable to the observed TM perforation alone. We aim with this study to better understand the effects of location and size of TM perforations on the sound transmitting properties of the middle ear.
The middle ear transfer function (METF) of six human temporal bones (TB; freshly frozen specimen of body donors) were compared before and after perforation of the TM at different locations (anterior or posterior lower quadrant) and of different sizes (1mm, ¼ of the TM, ½ of the TM, and full ablation). The
METF were correlated with a Finite Element (FE) model of the middle ear, in which similar alterations were simulated.
The measured and simulated FE model METFs exhibited frequency and perforation size dependent amplitude losses at all locations and severities. In direct comparison, posterior TM perforations affected the transmission properties to a larger degree than perforations of the anterior quadrant. This could possibly be caused by an asymmetry of the TM, where the malleus-incus complex rotates and results in larger deflections in the posterior TM half than in the anterior TM half. The FE model of the TM with a sealed cavity suggest that small perforations result in a decrease of TM rigidity and thus to an increase in oscillation amplitude of the TM, mostly above 1 kHz.
The location and size of TM perforations influence the METF in a reproducible way. Correlating our data with the FE model could help to better understand the pathologic mechanisms of middle-ear diseases. If small TM perforations with uncharacteristically significant HL are observed in daily clinical practice, additional middle ear pathologies should be considered. Further investigations on the loss of TM pretension due to perforations may be informative.
In order to evaluate the performance of different stapes prosthesis types, a coupled finite element (FE) model of human ear was developed. First, the middle-ear FE model was developed and validated using the middle-ear transfer function measurements available in literature including pathological cases. Then, the inner-ear FE model was developed and validated using tonotopy, impedance, and level of cochlea amplification curves from literature. Both models are based on pre-existing research with some improvements and were combined into one coupled FE model. The stapes in the coupled FE ear model was replaced with a model of a stapes prosthesis to create a reconstructed ear model that can be used to estimate how different types of protheses perform relative to each other as well as to the natural ear. This will help in designing of new innovative types of stapes prostheses or any other type of middle-ear prostheses as well as to improve the ones that are already available on the market.
Simulation models of the middle ear have rarely been used for diagnostic purposes due to their limited predictive ability with respect to pathologies. One big challenge is the large uncertainty and ambiguity in the choice of material parameters of the model.
Typically, the model parameters are determined by fitting simulation results to validation measurements. In a previous study, it was shown that fitting the model parameters of a finite-element model using the middle-ear transfer function and various other measurable output variables from normal ears alone is not sufficient to obtain a good predictive ability of the model on pathological middle-ear conditions. However, the inclusion of validation measurements on one pathological case resulted in a very good predictive ability also for other pathological cases. Although the found parameter set was plausible in all aspects, it was not yet possible to draw conclusions about the uniqueness and the accuracy or the uncertainty of the parameter set.
To answer these questions, statistical solution approaches are used in this study. Using the Monte Carlo method, a large number of plausible model data sets are generated that correctly represent the normal and pathological middle-ear characteristics in terms of various output variables like e.g., impedance, reflectance, umbo, and stapes transfer function. Subsequent principal component analyses (PCA) allow to draw conclusions about correlations, quantitative limits and statistical density of parameter values.
Furthermore, applying inverse PCA yields numerous plausible parameterizations of the middle-ear model, which can be used for data augmentation and training of a neural network which is capable of distinguishing between a normal middle ear and pathologies like otosclerosis, malleus fixation, and disarticulation based on objectively measured quantities like impedance, reflectance, and umbo velocity.
The hearing contact lens® (HCL) is a new type of hearing aid devices. One of its main components is a piezo-electric actuator. In order to evaluate and maximize the HCL’s performance, a model of the HCL coupled to the middle ear was developed using finite element approach. The model was validated step by step starting with the HCL only. To validate the HCL model, vibrational measurements on the HCL were performed using a Laser-Doppler-Vibrometer (LDV). Then, a silicone cap was placed onto the HCL to provide an interface between the HCL and the tympanic membrane of the middle-ear model and additional LDV measurements on temporal bones were performed to validate the coupled model. The coupled model was used to evaluate the equivalent sound pressure of the HCL. Moreover, a deeper insight was gained into the contact between the HCL and tympanic membrane and its effects on the HCL performance. The model can be used to investigate the sensitivity of geometrical and material parameters with respect to performance measures of the HCL and evaluate the feedback behavior.